| Customization: | Available |
|---|---|
| Application: | Electromagnetic Instrument |
| Power: | Electricity |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
| ☆ Wide Output |
|
|
| ☆ Wide Input |
|
|
| ☆ Multiple Output Mode |
|
|
|
|
| ☆ Multiple Output Mode |
|
|
|
☆ Graphics Features |
|
|
|
| S/N | Function declaration |
| 1 | Oscilloscope function,user can observe voltage and current waves.3 |
| 2 | Precise display phase for voltage and current. |
| 3 | Trend variation display. |
☆ Harmonics Function |
|
| S/N | Function declaration |
| 1 | Set amplitude of harmonic. |
| 2 | Set phase of harmonic(fundamental wave). |
| 3 | Choose 2nd~21st harmonics channel output. |
| 4 | Oscilloscope function, user can observe voltage and current waves. |
| 5 | Display frequency spectrum of harmonic by histogram.(fundament wave is 100%) |
| Voltage Range | Resolution | Stability ( %/min ) |
Accuracy ±(ppm of reading + ppm of range) [1] |
Max Burden (mA) |
||
| Class 0.05 | Class 0.02 | Class 0.05 | Class 0.02 | |||
| 57.7 V | 0.1 mV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 250 |
| 100 V | 1 mV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 150 |
| 220 V | 1 mV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 60 |
| 380 V | 1 mV | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 40 |
| Note [1] : (ppm = parts per million) (e.g.,10ppm = 0.001%). | ||||||
| Current Range | Resolution | Stability ( %/min ) |
Accuracy ±(ppm of reading + ppm of range) [1] |
Max Burden (V) |
||
| Class 0.05 | Class 0.02 | Class 0.05 | Class 0.02 | |||
| 200 mA | 1 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 20 |
| 500 mA | 1 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 10 |
| 1 A | 10 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 10 |
| 2 A | 10 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 2 |
| 5 A | 10 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 2 |
| 10 A | 0.1 mA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 2 |
| Symmetry | Voltage superior to 0.2%;Current superior to 0.5%;Phase superior to 0.5° |
| Frequency | Range:45 Hz~70 Hz; Adjustment fineness:0.001 Hz; Accuracy:±0.02 Hz (class 0.05), ±0.01 Hz(class 0.02) |
| Phase | Range:0.000 0°~359.999 9°; Adjustment fineness:0.001°; Accuracy:±0.02° (class 0.05), ±0.01°(class 0.02) |
| Harmonic | 2nd~21st harmonic; Amplitude 0~25% adjustable; Phase 0~359.99°adjustable |
| Current Range | Stability ( %/min ) |
Accuracy ( ± %*FS)[2] |
||
| Class 0.05 | Class 0.02 | Class 0.05 | Class 0.02 | |
| Active power |cosφ|≥0.5 | 0.01 | 0.005 | 0.05 | 0.02 |
| Reactive power |sinφ|≥0.5 | 0.02 | 0.01 | 0.1 | 0.05 |
| Apparent power | 0.02 | 0.01 | 0.1 | 0.05 |
| Power factor | 0.02 | 0.01 | 0.1 | 0.05 |
| Note [2] :FS= voltage range ×current range | ||||
| Type | Range | Resolution | Accuracy ±(ppm of reading + ppm of range) [1] |
|
| Direct Input | Clamp Current Input [3] | |||
| AC Voltage ACV |
57.7 V | 0.1 mV | 300 + 200 | -- |
| 100 V | 1 mV | 300 + 200 | -- | |
| 220 V | 1 mV | 300 + 200 | -- | |
| 380 V | 1 mV | 300 + 200 | -- | |
| AC Current ACI |
1 A | 10 μA | 300 + 200 | -- |
| 5 A | 10 μA | 300 + 200 | 0.2%*RG | |
| Note [3]:Clamp is option,if it be need, it must remark in order contract,Same as below. | ||||
| Type | Accuracy | |
| Direct Input | Clamp Current Input | |
| Active power | ± 0.05%*FS | ± 0.2%*FS |
| Reactive power | ± 0.1%*FS | ± 0.5%*FS |
| Apparent power | ± 0.1%*FS | ± 0.5%*FS |
| Power factor | ± 0.1% | ± 0.5% |
| Voltage Range | Resolution | Stability ( %/min ) |
Accuracy ±(ppm of reading + ppm of range) [1] |
Max Burden (mA) |
||
| Class 0.05 | Class 0.02 | Class 0.05 | Class 0.02 | |||
| 75 mV | 0.1 μV | 0.005 | 0.005 | 300 + 200 | 120 + 80 | 10 |
| 1 V | 10 μV | 0.005 | 0.005 | 300 + 200 | 120 + 80 | 10 |
| 10 V | 0.1 mV | 0.005 | 0.005 | 300 + 200 | 120 + 80 | 10 |
| 30 V | 0.1 mV | 0.005 | 0.005 | 300 + 200 | 120 + 80 | 500 |
| 100 V | 1 mV | 0.005 | 0.005 | 300 + 200 | 120 + 80 | 150 |
| 300 V | 1 mV | 0.005 | 0.005 | 300 + 200 | 120 + 80 | 50 |
| 600 V | 1 mV | 0.005 | 0.005 | 300 + 200 | 120 + 80 | 25 |
| Current Range | Resolution | Stability ( %/min ) |
Accuracy ±(ppm of reading + ppm of range) [1] |
Max Burden (V) |
||
| Class 0.05 | Class 0.02 | Class 0.05 | Class 0.02 | |||
| 1 mA | 10 nA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 3 |
| 5 mA | 10 nA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 3 |
| 20 mA | 100 nA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 3 |
| 100 mA | 1 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 3 |
| 1 A | 10 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 3 |
| 3 A | 10 μA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 3 |
| 10 A | 0.1 mA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 3 |
| Range | Measurement Range | Accuracy | Measurement Range of Ripple | Accuracy of Ripple |
| 1 V | ± ( 0~1.2 ) V | ± 0.01%*RG | 0~30 mV | ± 1 mV |
| 10 V | ± ( 0~12 ) V | ± 0.01%*RG | 0~300 mV | ± 10 mV |
| 2 mA | ±( 0~2.4) mA | ± 0.01%*RG | 0~60 μA | ± 2 μA |
| 20 mA | ± ( 0~24 ) mA | ± 0.01%*RG | 0~600 μA | ± 20 μA |
| Type | Accuracy | |
| Class 0.05 | Class 0.02 | |
| Active electrical energy | ± 0.1%*reading | ± 0.05%* reading |
| Reactive electrical energy | ± 0.2%* reading | ± 0.1%* reading |
| Power Supply | AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz |
| Temperature Performance | Working temperature:0°C~45°C; Storage temperature:-20°C~70°C |
| Humidity Performance | Working humidity:< 80% @ 30°C,< 70% @ 40°C,< 40% @ 50°C Storage humidity:(20%~80%) R·H,non-condensing |
| Interface | RS232 |