• TD1550 Verification Apparatus for DC Energy Meters
  • TD1550 Verification Apparatus for DC Energy Meters
  • TD1550 Verification Apparatus for DC Energy Meters
  • TD1550 Verification Apparatus for DC Energy Meters
  • TD1550 Verification Apparatus for DC Energy Meters
  • TD1550 Verification Apparatus for DC Energy Meters

TD1550 Verification Apparatus for DC Energy Meters

After-sales Service: Life Time
Warranty: 12 months Excellent After-sales Service
Power: Electricity
Customization:
Shipping Cost:

Estimated freight per unit.

about shipping cost and estimated delivery time.
Payment Method: visa mastercard discover JCB diners club american express T/T
  Initial Payment Full Payment
Currency: US$
Return&refunds: You can apply for a refund up to 30 days after receipt of the products.
From payment to delivery, we protect your trading.
TD1550 Verification Apparatus for DC Energy Meters pictures & photos
TD1550 Verification Apparatus for DC Energy Meters
US $63,340-92,500 / Set
Min. Order: 1 Set
Gold Member Since 2021

Suppliers with verified business licenses

Secured Trading Service
Manufacturer/Factory
  • Overview
  • Company Profile
Overview

Basic Info.

Model NO.
TD1550
Voltage Output
1mV~1150V
Current Output
1mA~600A
LCD Display
With LCD Display
Customized
Non-Customized
Transport Package
Package for Exporting
Trademark
TUNKIA
Origin
China
Production Capacity
1000

Packaging & Delivery

Package Size
350.00cm * 100.00cm * 130.00cm
Package Gross Weight
90.000kg

Product Description

TD1550 Verification Apparatus for DC Energy Meters
TD1550 Verification Apparatus for DC Energy Meters
* This is for your reference only,the configuration and details may vary depending on the application scenarios
1. Summary
TD1550 is a platform device that can be used for the verification of DC electric energy meter and DC shunt. It consists of DC voltage standard source, DC current standard source, DC small signal voltage standard source, DC voltmeter (option), multi position verification platform, verification software, etc. It can verify not only conventional DC electrical measuring instruments, but also DC shunt, direct and indirect connected DC energy meters.
 

2. Features

  1. Accuracy: class 0.02, class 0.05
  2. DC standard voltage source:10 mV~1150 V / 1550 V(option)
  3. DC standard current:1 mA~600 A
  4. DC small signal voltage source(four-line):10 μV~4.4 V
  5. Optional 3 or 6 meter position DC electric energy meter and DC shunt calibration stand
  6. Standard electric energy pulse input/output, used to detect the working error of electric energy meter
  7. Auxiliary power supply is used to supply power for electronic meters for measuring DC electrical energy(type A).
  8. Standard second pulse measurement is used to detect the daily timing error of the electric energy meter
  9. Mobile measurement and control station, which can observe or control the output through the LCD touch screen
  10. DC shunt manual/automatic crimping device (option), used for DC shunt verification.
 

3. Applications

Verification of DC electric energy meter
TD1550 Verification Apparatus for DC Energy Meters
  1. Application scenario 1: The maximum DC voltage output is 1550 V, meeting the requirements of 1.15 Un voltage test of 1 kV electric energy meter; The maximum DC current output is 600 A, meeting the requirements of 1.2 In current test of 100 A watt hour meter. With the power pulse measurement function, the test of direct connected DC energy meters.
  2. Application scenario 2:The instrument has 10 μ V~4.4 V four wire small signal voltage source, with 1550 V DC voltage source and electric energy pulse measurement function, can carry out relevant testing work of indirect connected DC energy meters.
  3. Application scenario 3:The instrument supports 600 A maximum current output option, with 1550 V DC voltage source and electric energy pulse measurement function, and can carry out relevant tests on the whole of indirect connected DC energy meters and DC shunt.


 
Verification of DC shunt (option)
TD1550 Verification Apparatus for DC Energy Meters
  1. DC high current standard source: the maximum output is 600 A, which can be used as the primary current input of the DC shunt.
  2. DC small signal voltmeter: support voltage input range is 10 μV~4.4 V, which can measure the secondary output voltage of the shunt
  3. DC shunt manual crimping device (option): manually compress the current terminal of the DC shunt and connect it with the current source to form a current loop.
  4. Verification and test of DC shunt (option): it can measure the resistance value and basic error of the shunt, and draw R(t) curve.


 
Calibration of DC electrical measuring instruments
TD1550 Verification Apparatus for DC Energy Meters
  1. Calibration of DC voltmeter and ammeter: the device has a built-in DC voltage standard source with an output range of 10 mV ~ 1550 V and a DC current standard source with an output range of 1 mA ~ 600 A. The voltage and current accuracy of the class 0.02 device reaches class 0.01, which is applicable to the calibration of DC voltmeter or ammeter of class 0.05 and below.
  2. Calibration of DC wattmeter: a virtual power standard source composed of independent DC voltage and DC current output. The accuracy of a class 0.02 device reaches class 0.02 , which is applicable to the calibration of DC power meters of class 0.1 and below.


 

4. Characteristics

Wide output / measurement range
TD1550 Verification Apparatus for DC Energy Meters
  1. The maximum output of its DC standard voltage source is 1550 V, and the maximum output of its DC standard current source is 600 A; The output range of the DC standard small signal voltage source used for the verification of indirect connected DC energy meters reaches ± (10 μ V~4.4 V)
  2. The maximum DC voltage measurement of the instrument can reach 4.4 V, and the maximum DC resistance measurement can reach 200 mΩ.
 
Clock check function  
TD1550 Verification Apparatus for DC Energy Meters  
  1. Built in clock module with precision up to 0.2 ppm, which can measure the second pulse signal for daily timing error test
  2. Built in GPS module and antenna can receive GPS standard clock signal.
 
Auxiliary power supply  
TD1550 Verification Apparatus for DC Energy Meters
  1. Each meter position has built-in auxiliary power supply, which can provide DC 24 V or AC 220 V power supply for electronic meters for measuring DC electrical energy (type A).
 

Four wire small signal voltage source
TD1550 Verification Apparatus for DC Energy Meters
  1. Four wire small signal voltage source:Precise output ± (10 μ V ~ 4.4 V) small signal voltage signal is used as the current input of indirect access DC watt hour meter (the specification of shunt can be set and displayed according to its primary current).
 
Professional test software
TD1550 Verification Apparatus for DC Energy Meters
  1. With RS232 communication interface, software functions can be customized according to customer needs
 

5. Specifications

5.1 DC Voltage Output

Range Resolution Short Term Stability
 (% / min)
Accuracy (k=2)
(ppm of reading + ppm of range) [1]
Maximum Burden Current (mA) Ripple  ( % )
Class 0.05 Class 0.02 Class 0.05 Class 0.02
100 mV 1 μV 0.005 3.5 μV 120 + 80 60 + 5 μV 200 < 1.0
300 mV 1 μV 0.005 0.002 120 + 80 60 + 40 200 < 1.0
1 V 10 μV 0.005 0.002 120 + 80 60 + 40 200 < 1.0
3 V 10 μV 0.005 0.002 120 + 80 60 + 40 200 < 1.0
10 V 0.1 mV 0.005 0.002 120 + 80 60 + 40 200 < 1.0
30 V 0.1 mV 0.005 0.002 120 + 80 60 + 40 200 < 1.0
100 V 1 mV 0.005 0.002 120 + 80 60 + 40 200 < 1.0
300 V 1 mV 0.005 0.002 120 + 80 60 + 40 60 < 1.0
600 V 1 mV 0.005 0.002 120 + 80 60 + 40 30 < 1.0
1000 V 10 mV 0.005 0.002 120 + 80 60 + 40 15 < 1.0
1500 V [2] 10 mV 0.005 0.002 120 + 80 60 + 40 10 < 1.0
Note [1] : (ppm = parts per million) (e.g.,10ppm = 0.001%).
[2]:1500V range is an option.
  1. Output range:10 mV~1150 V, 6-bit display
  2. 1500 V voltage output option can be selected to expand the voltage output range to 10 mV ~ 1550 V
  3. Protection function: Short circuit protection, overload protection
 

5.2 DC Current Output [3]

Range Resolution Short term stability
 (% / min)
Accuracy (k=2)
(ppm of reading + ppm of range) 
Maximum Burden Voltage (V) Ripple ( % )
Class 0.05 Class 0.02 Class 0.05 Class 0.02
10 mA 0.1 μA 0.005 0.003 120 + 80 60 + 40 12 < 0.5
20 mA 0.1 μA 0.005 0.003 120 + 80 60 + 40 12 < 0.5
50 mA 0.1 μA 0.005 0.003 120 + 80 60 + 40 12 < 0.5
100 mA 1 μA 0.005 0.003 120 + 80 60 + 40 12 < 0.5
200 mA 1 μA 0.005 0.003 120 + 80 60 + 40 12 < 0.5
500 mA 1 μA 0.005 0.003 120 + 80 60 + 40 3 < 0.5
1 A 10 μA 0.005 0.003 120 + 80 60 + 40 3 < 0.5
2 A 10 μA 0.005 0.003 120 + 80 60 + 40 3 < 0.5
5 A 10 μA 0.005 0.003 120 + 80 60 + 40 3 < 0.5
10 A 0.1 mA 0.005 0.003 120 + 80 60 + 40 3 < 0.5
20 A 0.1 mA 0.005 0.003 120 + 80 60 + 40 3 < 0.5
50 A 0.1 mA 0.005 0.003 120 + 80 60 + 40 3 < 1.0
100 A 1 mA 0.005 0.003 120 + 80 60 + 40 3 < 1.0
200 A 1 mA 0.005 0.003 120 + 80 60 + 40 3 < 1.0
500 A 1 mA 0.005 0.003 120 + 80 60 + 40 3 < 1.0
Note: [3] As the current input of the direct connected DC energy meters, or the current input of the external shunt of the indirect connected DC energy meters;
  1. Output range:1 mA~600 A,6-bit display
  2. Protection function: open circuit protection, overload protection
 

5.3 DC Small Signal Voltage Output[4]

Range Resolution Short Term Stability
(% / min)
Accuracy (k=2)
(ppm of reading + ppm of range)
Maximum Burden Current (mA) Ripple ( % )
1 mV 10 nV 0.1 μVrms 120 + 0.3 μV - < 0.2
3 mV 10 nV 0.3 μVrms 120 + 1 μV - < 0.2
10 mV 0.1 μV 1 μVrms 120 + 3 μV - < 0.2
30 mV 0.1 μV 0.002 120 + 80 - < 0.2
100 mV 1 μV 0.002 120 + 80 - < 0.2
300 mV 1 μV 0.002 120 + 80 ≤ 10 < 0.2
1 V 10 μV 0.002 120 + 80 ≤ 10 < 0.2
4 V 10 μV 0.002 120 + 80 ≤ 10 < 0.2
Note: [4] When the indirect connected DC energy meters is verified separately, it can be used as its current input (corresponding to the shunt voltage)
  1. Output range:± ( 10 μV~4.4 V ),6-bit display
  2. The specification and current voltage ratio of the shunt can be set, and the output value can be displayed according to the voltage or current converted by the shunts.

5.4 DC Shunts Verification

5.4.1 DC Voltage Measurement

Range Resolution Accuracy (k=2)
(ppm of reading + ppm of range or μV)
temperature coefficient
± ppm*RD /°C@15ºC~30ºC
1 mV 10 nV 120 + 2 μV 30
3 mV 10 nV 120 + 2 μV 30
10 mV 0.1 μV 120 + 2 μV 10
30 mV 0.1 μV 120 + 80 10
100 mV 1 μV 120 + 80 10
300 mV 1 μV 120 + 80 5
1 V 10 μV 120 + 80 5
4 V 10 μV 120 + 80 5
  1. Output range:± ( 10 μV~4.4 V ),6-bit display,Manual/automatic range switching
  2. Input resistance:>1GΩ;Input protection:±50Vpk,continuity.


5.4.2 DC Resistance Measurement

Test current Voltage Range Measuring Range Of Resistance Accuracy (k=2)
 @ Full Scale Voltage Value[5]
Class 0.05 Class 0.02
5 A 1 mV 20.0000 μΩ~200.000 μΩ 0.232% 0.222%
10 mV 200.000 μΩ~2.00000 mΩ 0.052% 0.042%
100 mV 2.000 00 mΩ~20.000 0 mΩ 0.04% 0.03%
1 V 20.000 0 mΩ~200.000 mΩ 0.04% 0.03%
10 A 1 mV 10.000 0 μΩ~100.000 μΩ 0.232% 0.222%
10 mV 100.000 μΩ~1.000 00 mΩ 0.052% 0.042%
100 mV 1.000 00 mΩ~10.000 0 mΩ 0.04% 0.03%
1 V 10.000 0 mΩ~100.000 mΩ 0.04% 0.03%
20 A 1 mV 5.000 00 μΩ~50.000 0 μΩ 0.232% 0.222%
10 mV 50.000 0 μΩ~500.000 μΩ 0.052% 0.042%
100 mV 500.000 μΩ~5.000 00 mΩ 0.04% 0.03%
1 V 5.000 00 mΩ~50.000 0 mΩ 0.04% 0.03%
50 A 1 mV 2.000 00 μΩ~20.000 0 μΩ 0.232% 0.222%
10 mV 20.000 0 μΩ~200.000 μΩ 0.052% 0.042%
100 mV 200.000 μΩ~2.000 00 mΩ 0.04% 0.03%
1 V 2.000 00 mΩ~20.000 0 mΩ 0.04% 0.03%
100 A 1 mV 1.000 00 μΩ~10.000 0 μΩ 0.232% 0.222%
10 mV 10.000 0 μΩ~100.000 μΩ 0.052% 0.042%
100 mV 100.000 μΩ~1.000 00 mΩ 0.04% 0.03%
1 V 1.000 00 mΩ~10.000 0 mΩ 0.04% 0.03%
200 A 1 mV 500.000 nΩ~5.000 00 μΩ 0.232% 0.222%
10 mV 5.000 00 μΩ~50.000 0 μΩ 0.052% 0.042%
100 mV 50.000 0 μΩ~500.000 μΩ 0.04% 0.03%
1 V 500.000 μΩ~5.000 00 mΩ 0.04% 0.03%
500 A 1 mV 200.000 nΩ~2.000 00 μΩ 0.232% 0.222%
10 mV 2.000 00 μΩ~20.000 0 μΩ 0.052% 0.042%
100 mV 20.000 0 μΩ~200.000 μΩ 0.04% 0.03%
1 V 200.000 μΩ~2.000 00 mΩ 0.04% 0.03%
 
  1. Measuring range of resistance:200.000 nΩ~4.00000 Ω,6-bit display,Minimum resolution:1 pΩ
  2. The current test point can be set through the number keys, and the voltage range: manual/automatic switching
  3. After loading the test current, the R (t) change curve can be tested
 

5.5 Daily timing error

  1. Annual measurement uncertainty of device standard clock:2×10-7
  2. Daily timing error supports two error display modes: s/d and ppm/ppb
 

5.6 DC Electric energy index

DC Power/Electric Energy Range Combination of DC voltage and DC current (or DC small signal voltage)
Accuracy Uncertainty of voltage measurement+uncertainty of current measurement (or uncertainty of DC small signal voltage measurement)
Electric energy pulse Standard Electric Energy Pulse Output High frequency full range value corresponds to 60 kHz;
Low frequency full range value corresponds to 6 Hz;
 
Pulse Output Frequency Output frequency jitter time<10 μ s
Standard Electric Energy Pulse Input Signal amplitude is 5V, TTL level;
Able to receive high-frequency pulses ≤ 150 kHz;
FH=60 kHz × Power value ÷ voltage range value ÷ current range value
FL=60 kHz × Power value ÷ voltage range value ÷ current range value ÷ 10000
Electric Energy Error Display Automatic display, resolution 0.0001%
  1.  

5.7 Auxiliary power supply

DC Power Output OutputVoltage 24 V
Maximum Burden Current 120 mA / Epitope
Protection Function Replaceable fuse protection
AC Power Output Output Voltage 220 V
Maximum Burden Current 20 mA / Epitope
Protection Function Replaceable fuse protection
 

6. Ordering Information

TD1550 Verification Apparatus for DC Energy Meters

7. General Specifications

Power Supply AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz;
Temperature Performance Working temperature:15 °C~30 °C;
Storage temperature:-20 °C ~ 70 °C;
Humidity Performance Working humidity:( 20 % ~ 50% ) R·H,No condensation;
Storage humidity:( 15 % ~ 80% ) R·H,No condensation;
Interface RS232

Company Profile
 

TD1550 Verification Apparatus for DC Energy Meters

TD1550 Verification Apparatus for DC Energy Meters

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now

You Might Also Like

From payment to delivery, we protect your trading.
TD1550 Verification Apparatus for DC Energy Meters pictures & photos
TD1550 Verification Apparatus for DC Energy Meters
US $63,340-92,500 / Set
Min. Order: 1 Set
Gold Member Since 2021

Suppliers with verified business licenses

Secured Trading Service
Manufacturer/Factory
Plant Area
8000 square meters
Management System Certification
ISO9001:2015, ISO14001:2015, ISO45001:2016, Intellectual Property Management System Certificate, ISO 27001:2013