☆ Calibrate High-current DC Meter |
|
☆ Verification of DC Shunt |
|
☆ High accuracy and stability of current output |
Precise integrated circuit devices are used to ensure the long-term accuracy of the system.
Adopt the design principle of closed-loop negative feedback system to ensure the stability of output.
|
☆ Convenience of Operation |
Digital programmable keys: It can realize fixed-point output, rotary encoder, step adjustment and multiple input modes.
|
☆ Multiple Output Methods |
The instrument has a "fixed-point output" mode. The required output value can be directly set by pressing the number keys on the console or clicking the touch screen, and the instrument will automatically switch to the optimal range output;
In particular, when setting the current, input the current value directly.
|
|
The key operation area is equipped with a "rotary knob", which can increase or decrease the value output by rotating clockwise or counterclockwise.
|
☆ Proportional Output |
When calibrating electrical measuring instruments, it is usually necessary to select calibration points according to the proportion of each range of the meter being tested;
The user can select the calibration point of the tested meter through the "percentage verification point" on the touch screen of the instrument.
|
☆ Professional Test Software (Option) |
The device has RS232 communication interfaces, and its software functions can be customized according to customer needs.
|
Range | Short-term Stability (% / min) | Accuracy ( ± ppm of reading + ppm of range) [1] |
Maximum load (V) | ||
Class 0.05 | Class 0.02 | Class 0.05 | Class 0.02 | ||
5 A | 0.005 | 0.003 | 120 + 80 | 60 + 40 | 3.5 |
10 A | 0.005 | 0.003 | 120 + 80 | 60 + 40 | 3.5 |
20 A | 0.005 | 0.003 | 120 + 80 | 60 + 40 | 3.5 |
50 A | 0.005 | 0.003 | 120 + 80 | 60 + 40 | 3.5 |
100 A | 0.005 | 0.003 | 120 + 80 | 60 + 40 | 3.5 |
200 A | 0.005 | 0.003 | 120 + 80 | 60 + 40 | 3.5 |
500 A | 0.005 | 0.003 | 120 + 80 | 60 + 40 | 3.5 |
Note [1] : (ppm = parts per million) (e.g., 10ppm = 0.001%). |
Range | Accuracy ( ± ppm of reading + ppm of range) |
Temperature coefficient @ (15~30) °C (ppm*RG/°C) |
|||
Class 0.05 | Class 0.02 | Class 0.05 | Class 0.02 | ||
1 mV | 150 + 1 | 80 + 0.5 | <30 | <15 | |
10 mV | 150 + 3 | 80 + 1.5 | <10 | <5 | |
100 mV | 150 + 10 | 80 + 5 | <10 | <5 | |
1 V | 150 + 20 | 80 + 10 | <5 | <2 | |
10 V | 150 + 100 | 80 + 50 | <5 | <2 |
Test Current | Voltage Range | Measuring Range of Resistance | Accuracy @ Full scale voltage value |
|
0.05 | 0.02 | |||
5 A | (0.1~1) mV | 20 μΩ ≤ R ≤ 200 μΩ | 0.135% | 0.068% |
(1~10) mV | 200 μΩ < R ≤ 2 mΩ | 0.065% | 0.033% | |
(10~100) mV | 2 mΩ < R ≤ 20 mΩ | 0.045% | 0.023% | |
(0.1~1) V | 20 mΩ < R ≤ 200 mΩ | 0.037% | 0.019% | |
(1~3) V | 200 mΩ < R ≤600 mΩ | 0.036% | 0.019% | |
10 A | (0.1~1) mV | 10 μΩ ≤ R ≤ 100 μΩ | 0.135% | 0.068% |
(1~10) mV | 100 μΩ < R ≤ 1 mΩ | 0.065% | 0.033% | |
(10~100) mV | 1 mΩ < R ≤ 10 mΩ | 0.045% | 0.023% | |
(0.1~1) V | 10 mΩ < R ≤ 100 mΩ | 0.037% | 0.019% | |
(1~3) V | 100 mΩ < R ≤300 mΩ | 0.036% | 0.019% | |
20 A | (0.1~1) mV | 5 μΩ ≤ R ≤ 50 μΩ | 0.135% | 0.068% |
(1~10) mV | 50 μΩ < R ≤ 500 μΩ | 0.065% | 0.033% | |
(10~100) mV | 500 μΩ < R ≤ 5 mΩ | 0.045% | 0.023% | |
(0.1~1) V | 5 mΩ < R ≤ 50 mΩ | 0.037% | 0.019% | |
(1~3) V | 50 mΩ < R ≤ 150 mΩ | 0.036% | 0.019% | |
50 A | (0.1~1) mV | 2 μΩ ≤ R ≤ 20 μΩ | 0.135% | 0.068% |
(1~10) mV | 20 μΩ < R ≤ 200 μΩ | 0.065% | 0.033% | |
(10~100) mV | 200 μΩ < R ≤ 2 mΩ | 0.045% | 0.023% | |
(0.1~1) V | 2 mΩ < R ≤ 20 mΩ | 0.037% | 0.019% | |
(1~3) V | 20 mΩ < R ≤ 60 mΩ | 0.036% | 0.019% | |
100 A | (0.1~1) mV | 1 μΩ ≤ R ≤ 10 μΩ | 0.135% | 0.068% |
(1~10) mV | 10 μΩ < R ≤ 100 μΩ | 0.065% | 0.033% | |
(10~100) mV | 100 μΩ < R ≤ 1 mΩ | 0.045% | 0.023% | |
(0.1~1) V | 1 mΩ < R ≤ 10 mΩ | 0.037% | 0.019% | |
(1~3) V | 10 mΩ < R ≤ 30 mΩ | 0.036% | 0.019% | |
200 A | (0.1~1) mV | 500 nΩ ≤ R ≤ 5 μΩ | 0.135% | 0.068% |
(1~10) mV | 5 μΩ < R ≤ 50 μΩ | 0.065% | 0.033% | |
(10~100) mV | 50 μΩ < R ≤ 500 μΩ | 0.045% | 0.023% | |
(0.1~1) V | 500 μΩ < R ≤ 5 mΩ | 0.037% | 0.019% | |
(1~3) V | 5 mΩ< R ≤ 15 mΩ | 0.036% | 0.019% | |
500 A | (0.1~1) mV | 200 nΩ ≤ R ≤ 2 μΩ | 0.135% | 0.068% |
(1~10) mV | 2 μΩ < R ≤ 20 μΩ | 0.065% | 0.033% | |
(10~100) mV | 20 μΩ < R ≤ 200 μΩ | 0.045% | 0.023% | |
(0.1~1) V | 200 μΩ < R ≤ 2 mΩ | 0.037% | 0.019% | |
(1~3) V | 2 mΩ< R ≤ 6 mΩ | 0.036% | 0.019% |
Power supply | AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz; |
Warm up time | 2h; |
Maximum power consumption | 4.8 kVA |
Sampling rate | 0.5 sps |
Temperature performance | Operating temperature: 15 °C~30 °C; Storage temperature: -20 °C ~ 70 °C; |
Humidity performance | Operating humidity: ( 20 % ~ 50% ) R·H, No condensation; Storage humidity: ( 15 % ~ 80% ) R·H, No condensation; |
Altitude | < 3000 m |
Weight | About 49.5 kg |
Communication interface | RS 232 |
Unit size | 623 mm (W) × 548 mm (D) × 260 mm (H) (Excluding handle and support foot) |