Product Highlights

  • High Precision Calibration: Accuracy classes 0.02, 0.05, and 0.1 for reliable earth continuity tester calibration.
  • Advanced Current Measurement: Measures AC/DC current from 0.5 A to 60 A with optional high-range configurations.
  • Harmonic & Ripple Analysis: Supports 2nd-32nd harmonic measurement and DC ripple test with up to 1 kHz bandwidth.
  • Four-Terminal Connection: Eliminates lead resistance errors for precise low-resistance measurements.
  • LCD Touch Screen Interface: Intuitive operation with real-time waveform display and data statistics functions.
  • Comprehensive Data Statistics: Records Max, Min, Avg, Span, Stab, and deviation for detailed test analysis.
  • Lifetime After-Sales Service: Provides lifetime technical support and service for the TD1250 calibration device.

Product Images of Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester

Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester pictures & photos
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester pictures & photos
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester pictures & photos
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester pictures & photos
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester pictures & photos
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester pictures & photos
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
  • Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
    Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
  • Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
    Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
  • Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
    Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
  • Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
    Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
  • Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
    Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
  • Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
    Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester

Find Similar Products

Product Description

TD1250 Testing Apparatus for Earth Continuity Testers
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester

1. Summary

TD1250 is a precision AC/DC artificial resistance, intergraded a AC/DC ammeter. Applied for calibrating earth continuity tester and other AC/DC low resistance instruments. It can calibrate max permissible error, alarm setting error, test current setting error etc.
 

2. Features

  1. DC resistance simulation:200 μΩ (100 μΩ optional)~20 Ω.
  2. AC resistance simulation:500 μΩ (300 μΩ optional)~15 Ω.
  3. AC/DC current measurement:0.5 A~36 A(60A option).
  4. Accuracy class:0.02/ 0.05/ 0.1.
  5. Harmonic and ripple function.
  6. Current waveform display function.
  7. With data statistics and analysis, stability test function.
  8. LCD touch screen.
  9. Setting output values by keys.

3. Applications

Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester

 

4. Characteristics

☆ Resistance Output
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
  1. Using four-terminals connections, can effective eliminate the measurement error introduced by the lead resistance.


 
☆ Multiple Output Mode
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
  1. "Direct output"mode,User can set output value by physical key or touch screen.
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
  1. "Rotary Knob"mode,User can setting in clockwise direction or anticlockwise direction.



 
☆ Multiple Output Mode
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
  1. Touch "Calibration point"of screen for "% setting".


 

☆ AC+DC Measurement
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
  1. Support DC/AC)measurement mode. AC frequency range:45 Hz ~ 65 Hz.
  2. Distortion and ripple test,max bandwidth of DC ripple reaches 1 kHz. Supports 2nd~32nd harmonics.



 
☆ DC Characteristic Analysis
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
Serial Number Explanation
  Current overshoot : The maximum value of the test current is displayed in DC mode, which is used to calculate the operating current overshoot;
  Waveform display:With oscilloscope function ,can real time display of measured current waveform.
  Ripple measurement:the instrument directly displays the ripple measurement results on the screen for users to read.



 
☆ AC Characteristic Analysis
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
Serial Number Explanation
  Harmonic distortion:Display THD/T(harmonic vs. full wave),THD/F(harmonic vs. fundamental). Easy for users to real intuitively.
  Harmonic data:Display the amplitude (RMS), content(%), phase of the 2nd~32nd harmonic, facilitate user analysis.
  DC component:Display the DC component in the AC current signal. Easy for users to real intuitively.
  Waveform display:with waveform display function, real time display of measured current waveform.




 
☆ Statistical Analysis
Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
Serial Number Function Description
  Scoring cycle:0.5 s ~ 20 s integration period can be set(1s by default); it is convenient for users to set sampling data according to actual needs.
  Statistical analysis:Real time record and statistics of current test data, including Max, Min, Avg, Span, Stab, S.dev, Time, test times N,etc; Easy for users to real intuitively.


 

5. Specifications

5.1 DC Current Measurement 

Range Resolution Accuracy
±(ppm of reading + ppm of range) [1]
Class 0.1 Class 0.05 Class 0.02
1 A 10 μA 0.05 + 1.5 0.02 + 1.0 0.01 + 0.5
2 A 10 μA 0.05 + 1.5 0.02 + 1.0 0.01 + 0.5
5 A 10 μA 0.05 + 1.5 0.02 + 1.0 0.01 + 0.5
10 A 100 μA 0.05 + 1.5 0.02 + 1.0 0.01 + 0.5
20 A 100 μA 0.05 + 2.5 0.02 + 2.0 0.01 + 1.0
30 A 100 μA 0.05 + 5.0 0.02 + 3.0 0.01 + 2.0
50 A [2] 100 μA 0.05 + 5.0 0.02 + 3.0 0.01 + 2.0
Note [1] : (ppm = parts per million) (e.g.,10ppm = 0.001%). Note [2]: 50 A Range is optional Accessories
  1. Measuring range(TD1250-30A specification, max range:30 A):0.5 A~36 A.
  2. Measuring range(TD1250-50A specification, max range: 50 A):0.5 A~60 A.
  3. Ripple frequency bandwidth: 1 Hz ~ 1 kHz, accuracy: ±0.02%*RG,effective value.
 

5.2 AC Current Measurement

Range Resolution Accuracy
±(ppm of reading + ppm of range) [1]
Class 0.1 Class 0.05 Class 0.02
1 A 10 μA 0.10 + 3.0 0.05 + 2.0 0.02 + 1.0
2 A 10 μA 0.10 + 3.0 0.05 + 2.0 0.02 + 1.0
5 A 10 μA 0.10 + 3.0 0.05 + 2.0 0.02 + 1.0
10 A 100 μA 0.10 + 3.0 0.05 + 2.0 0.02 + 1.0
20 A 100 μA 0.10 + 5.0 0.05 + 3.0 0.02 + 2.0
30 A 100 μA 0.10 + 10 0.05 + 5.0 0.02 + 5.0
50 A[2] 100 μA 0.10 + 10 0.05 + 5.0 0.02 + 5.0
  1. Measuring range(TD1250-30A specification, max range:30 A):0.5 A~36 A.
  2. Measuring range(TD1250-50A specification, max range: 50 A):0.5 A~60 A.
  3. AC frequency: 45 Hz ~ 65 Hz, accuracy: ±0.01 Hz, harmonic measurement: 2nd~32nd.
 

5.3 DC Simulated Resistance

Current Range Setting Range of Resistance Value R Adjustment Fineness Accuracy
±(A%*reading+R0)
Class 0.1 Class 0.05 Class 0.02
50 A [2] 100 μΩ ≤ R ≤ 4 mΩ 20 nΩ 0.1 + 3 μΩ 0.08 + 1.5 μΩ 0.04 + 0.8 μΩ
4 mΩ < R ≤ 40 mΩ 200 nΩ 0.1 + 3 μΩ 0.06 + 1.5 μΩ 0.03 + 0.8 μΩ
40 mΩ < R ≤ 400 mΩ 2 μΩ 0.1 0.05 0.02
30A 200 μΩ ≤ R ≤ 6 mΩ 30 nΩ 0.1 + 5 μΩ 0.08 + 3 μΩ 0.04 + 1.5 μΩ
6 mΩ < R ≤ 60 mΩ 300 nΩ 0.1 + 5 μΩ 0.06 + 3 μΩ 0.03 + 1.5 μΩ
60 mΩ < R ≤ 600 mΩ 3 μΩ 0.1 0.05 0.02
20A 400 μΩ ≤ R ≤ 10 mΩ 100 nΩ 0.1 + 6 μΩ 0.08 + 5 μΩ 0.04 + 2.5 μΩ
10 mΩ < R ≤ 100 mΩ 1 μΩ 0.1 + 6 μΩ 0.06 + 5 μΩ 0.03 + 2.5 μΩ
100 mΩ ≤ R ≤ 1 Ω 10 μΩ 0.1 0.05 0.02
10A 800 μΩ ≤ R ≤ 20 mΩ 100 nΩ 0.1 + 15 μΩ 0.08 + 6 μΩ 0.04 + 3 μΩ
20 mΩ < R ≤ 200 mΩ 1 μΩ 0.1 + 15 μΩ 0.06 + 6 μΩ 0.03 + 3 μΩ
200 mΩ < R ≤ 2 Ω 10 μΩ 0.1 0.05 0.02
5A 1.6 mΩ ≤ R ≤ 40 mΩ 200 nΩ 0.15 + 30 μΩ 0.09 + 15 μΩ 0.045 + 6 μΩ
40 mΩ < R ≤ 400 mΩ 2 μΩ 0.12 + 30 μΩ 0.07 + 15 μΩ 0.035 + 6 μΩ
400 mΩ < R ≤ 4 Ω 20 μΩ 0.1 0.05 0.025
2A 4 mΩ ≤ R ≤ 100 mΩ 1 μΩ 0.20 + 50 μΩ 0.10 + 30 μΩ 0.055 + 15 μΩ
100 mΩ < R ≤ 1 Ω 10 μΩ 0.20 + 50 μΩ 0.10 + 30 μΩ 0.045 + 15 μΩ
1 Ω < R ≤ 10 Ω 100 μΩ 0.1 0.05 0.035
1A 8 mΩ ≤ R ≤ 200 mΩ 1 μΩ 0.30 + 90 μΩ 0.15 + 50 μΩ 0.08 + 30 μΩ
200 mΩ < R ≤ 2 Ω 10 μΩ 0.30 + 90 μΩ 0.15 + 50 μΩ 0.07 + 30 μΩ
2 Ω < R ≤ 20 Ω 100 μΩ 0.2 0.1 0.06
Note [2]: 50 A Range is optional Accessories
  1. Simulated resistance range(TD1250-30A specification, maximum range: 30 A):200 μΩ~20 Ω
Simulated resistance range(TD1250-50A specification, maximum range: 50 A):100 μΩ~20 Ω
  1. Display Digits:6-digit decimal display, set the resistance value through the number keys.

5.4 AC Simulated Resistance

Current Range Setting Range of Resistance Value R Adjustment Fineness Accuracy
±(A%*reading+R0)
Class 0.1 Class 0.05 Class 0.02
50 A [2] 300 μΩ ≤ R ≤ 3 mΩ 20 nΩ 0.3 + 5 μΩ 0.15 + 3 μΩ 0.08 + 1.5 μΩ
3 mΩ < R ≤ 30 mΩ 200 nΩ 0.24 + 5 μΩ 0.12 + 3 μΩ 0.06 + 1.5 μΩ
30 mΩ < R ≤ 300 mΩ 2 μΩ 0.2 0.1 0.04
30A 500 μΩ ≤ R ≤ 5 mΩ 30 nΩ 0.3 + 8 μΩ 0.15 + 5 μΩ 0.08 + 2.5 μΩ
5 mΩ < R ≤ 50 mΩ 300 nΩ 0.24 + 8 μΩ 0.12 + 5 μΩ 0.06 + 2.5 μΩ
50 mΩ < R ≤ 500 mΩ 3 μΩ 0.2 0.1 0.04
20A 750 μΩ ≤ R ≤ 7.5 mΩ 50 nΩ 0.3 + 10 μΩ 0.15 + 6 μΩ 0.08 + 4 μΩ
7.5 mΩ < R ≤ 75 mΩ 500 nΩ 0.24 + 10 μΩ 0.12 + 6 μΩ 0.06 + 4 μΩ
75 mΩ ≤ R ≤ 750 mΩ 5 μΩ 0.2 0.10 0.04
10A 1.5 mΩ < R ≤ 15 mΩ 100 nΩ 0.3 + 20 μΩ 0.15 + 12 μΩ 0.08 + 8 μΩ
15 mΩ < R ≤ 150 mΩ 1 μΩ 0.24 + 20 μΩ 0.12 + 12 μΩ 0.06 + 8 μΩ
150 mΩ < R ≤ 1.5 Ω 10 μΩ 0.2 0.1 0.04
5A 3 mΩ ≤ R ≤ 30 mΩ 200 nΩ 0.3 + 20 μΩ 0.2 + 15 μΩ 0.1 + 12 μΩ
30 mΩ < R ≤ 300 mΩ 2 μΩ 0.3 + 20 μΩ 0.15 + 15 μΩ 0.07 + 12 μΩ
300 mΩ < R ≤ 3 Ω 20 μΩ 0.2 0.1 0.05
2A 7.5 mΩ ≤ R ≤ 75 mΩ 5 μΩ 0.4 + 80 μΩ 0.2 + 50 μΩ 0.11 + 30 μΩ
75 mΩ < R ≤ 750 mΩ 50 μΩ 0.4 + 80 μΩ 0.2 + 50 μΩ 0.09 + 30 μΩ
750 mΩ < R ≤ 7.5 Ω 500 μΩ 0.2 0.14 0.07
1A 15 mΩ ≤ R ≤ 150 mΩ 1 μΩ 0.6 + 150 μΩ 0.3 + 90 μΩ 0.16 + 60 μΩ
150 mΩ < R ≤ 1.5 Ω 10 μΩ 0.6 + 150 μΩ 0.3 + 90 μΩ 0.14 + 60 μΩ
1.5 Ω < R ≤ 15 Ω 100 μΩ 0.4 0.2 0.12
Note [2]: 50 A Range is optional Accessories
  1. Simulated resistance range(TD1250-30A specification, maximum range: 30 A):500 μΩ~15 Ω.
Simulated resistance range(TD1250-50A specification, maximum range: 50 A):300 μΩ~15 Ω.
  1. Display Digits:6-digit decimal display, set the resistance value through the number keys.
  2. AC frequency: 45 Hz ~ 65 Hz, accuracy: ±0.01 Hz.

6. General Specifications

Power Supply AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz
Temperature Performance Working temperature:0°C~45°C;
Storage temperature:-20°C~70°C
Humidity Performance Working humidity:< 80% @ 30°C,< 70% @ 40°C,< 40% @ 50°C
Storage humidity:(20%~80%) R·H,non-condensing
Interface RS232
 

7. Ordering Information

Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester
Company Profile
 

Tunkia TD1250 Earth Continuity Testers Calibration Device for Earth Continuty Resistance Tester


About the Supplier

This supplier has been audited by SGS. Verify the report on the official SGS website with ID "QIP-ASI2191183". Items marked "verified" are certified.

Watch Video
Watch Video
Tunkia Co., Ltd.
Tunkia Co., Ltd.
Tunkia Co., Ltd.
Tunkia Co., Ltd.
Tunkia Co., Ltd.
Tunkia Co., Ltd.
Tunkia Co., Ltd.
Tunkia Co., Ltd.

About Our Factory & Business Background

verified Business Type
Manufacturer/Factory
Address
No. 16 Panpan Road, Changsha, Hunan, China
verified Plant Area
8000 square meters
verified Number of Employees
363
verified Registered Capital
25,000,000 RMB
verified Terms of Payment
LC, T/T, D/P, PayPal, Western Union
verified International Commercial Terms(Incoterms)
FOB, CIF, CFR, EXW
verified Average Lead Time
Peak Season Lead Time: one month
Off Season Lead Time: one month
verified Nearest Port
Changsha Port

Our Production Capability & Technical Expertise

verified Main Products
Magnetic Measurement, Electrical Measurement, Gaussmeter, Multifunction Calibrator
verified Production Machines
SMT Machine, Printing Machine, Assembly & Debugging Line, Automatic Online Function Test (EPS&SST)
verified ODM Service Available
Yes
verified OEM Service Available
Yes
verified R&D Engineers
120 people
verified Own Brand
Yes

Our Industry Experience & Global Business Record

verified Year of Establishment
2006-06-16
Export Year
2021-10-11
verified Main Markets
Europe, Southeast Asia/ Mideast, East Asia(Japan/ South Korea), Domestic
verified Number of Foreign Trading Staff
2
verified Overseas Agent/Branch
No

FAQ

What is the warranty period for the TD1250?
The device comes with a 12-month warranty and lifetime after-sales service.
What accuracy classes are available?
The TD1250 offers accuracy classes of 0.02, 0.05, and 0.1.
Can the device measure harmonics?
Yes, it supports 2nd to 32nd harmonic measurement and displays THD/T and THD/F.
What is the maximum current output?
The standard range is up to 36 A, with an optional 50 A range reaching up to 60 A.
Does it support AC and DC measurements?
Yes, it supports both AC (45-65 Hz) and DC measurement modes with ripple analysis.
How is the resistance value set?
Users can set output values via physical keys or the LCD touch screen.
Is the device customizable?
No, the product is listed as Non-Customized.

Send your message to this supplier

Contact Information
Recipient
Message

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now