Output Range | 0~50%*range value (0 V~504 V) |
Display Digit | 7-digit display |
Short Term Stability | 0.003%/min |
Annual Accuracy (tcal±5ºC) | ± (0.009%*reading value+0.01%*range) |
Maximum Load Current (Peak) | 1 A (U ≤ 600V range); 71 mA (1000V range) |
Output Range | 1 V~1008 V |
Output Frequency Range | 16 Hz~850 Hz |
Display Digits | 7-digit display |
Short Term Stability | 0.0015%/min |
Annual Accuracy (tcal±5ºC) | 16 Hz≤F<45 Hz: ± (0.006%* reading value +0.001%* range) 45 Hz≤F≤65 Hz: ± (0.004%* reading value +0.001%* range) 65 Hz<F≤850 Hz: ± (0.006%* reading value +0.001%* range) |
Maximum Load Current (Peak) | 1 A (U ≤ 600V range); 71 mA (1000V range) |
Maximum Harmonic Distortion | 0.016%* set value or 0.003%* range value, use the larger value of the two |
Output Range | 0~30%* range value |
Harmonic Order | 2nd~99th @50Hz / 60Hz |
Short Term Stability | 0.002%/min |
Annual Accuracy (tcal±5ºC) | 16 Hz≤F≤850 Hz:±(0.006%* reading value +0.002%* range) 850 Hz<F≤6 kHz:±(0.045%* reading value +0.0024%* range) |
Maximum Load Current (Peak) | 1 A(U ≤ 600V range); 71 mA(1000V range) |
Maximum Harmonic Distortion | 16 Hz≤F≤850 Hz: 0.016%* set value or 0.003%* range value, use the larger value of the two 850 Hz<F≤6 kHz: 0.25%* set value or 0.015%* range value, use the larger value of the two |
Output Range | 0 A~24 A |
Display Digits | 7-digit display |
Short Term Stability | 0.003%/min |
Annual Accuracy (tcal±5ºC) | ± (0.009%*reading value +0.01%*range) |
Maximum Compliance Voltage (Peak) | 14 Vpk@ (0.25A~10A range) |
Output Range | 0.1 A~80 A |
Output Frequency Range | 16 Hz~850 Hz |
Display Digits | 7-digit display |
Short Term Stability | 0.002%/min |
Annual Accuracy | 16 Hz≤F<45 Hz: ± (0.006%* reading value+0.002%* range) 45 Hz≤F≤65 Hz: ± (0.005%* reading value+0.001%* range) 65 Hz<F≤850 Hz: ± (0.006%* reading value+0.002%* range) |
Maximum Compliance Voltage (Peak) | 14 Vpk@(0.25A~10A range) |
Maximum Harmonic Distortion | 0.01%* set value or 0.003%* range value, use the larger value of the two |
Output Range | 0.1 A~24 A |
Harmonic Order | 2nd~99th @50Hz / 60Hz |
Short Term Stability | 0.002%/min |
Annual Accuracy (tcal±5ºC) | 16 Hz≤F≤850 Hz: ± (0.006%* reading value +0.002%* range) 850 Hz<F≤6 kHz: ± (0.04%* reading value +0.002%* range) |
Maximum Compliance Voltage (Peak) | 14 Vpk@(0.25A~10A range) |
Maximum Harmonic Distortion | 16 Hz≤F≤850 Hz: 0.016%* set value or 0.003% range value, use the larger value of the two 850 Hz<F≤6 kHz: 0.25%* set value or 0.015% range value, use the larger value of the two |
Output Range | 0~50%* range value (0 V~5 V) |
Display Digits | 7-digit display |
Short Term Stability | 0.003%/min |
Annual Accuracy (tcal±5ºC) | ±(0.009%* reading value +0.014%* range) |
Output Range | 0.05 V~10 V |
Output Frequency Range | 16 Hz~850 Hz |
Display Digits | 7-digit display |
Short Term Stability | 0.002%/min |
Annual Accuracy (tcal±5ºC) | 45 Hz≤F≤65 Hz: ± (0.005%* reading value +0.002%* range) 16 Hz<F≤850 Hz: ± (0.006%* reading value +0.002%* range) |
Maximum Harmonic Distortion | 0.010%* set value or 0.001%* range value, use the larger value of the two |
Output Range | 0~3V |
Harmonic Order | 2nd~99th @50Hz / 60Hz |
Short Term Stability | 0.002%/min |
Annual Accuracy (tcal±5ºC) | 16 Hz≤F≤850 Hz: ± (0.0065%* reading value +0.0020% range) 850 Hz<F≤6 kHz: ± (0.04%* reading value +0.0030%* range) |
Maximum Harmonic Distortion | 16 Hz≤F≤850 Hz: 0.010%* set value or 0.001%* range value, use the larger value of the two 850 Hz<F≤6 kHz: 0.100%* set value or 0.01%* range value, use the larger value of the two |
Frequency | Annual Accuracy (tcal ±5ºC) | |||
>40% Range Voltage and Current Components | 0.5%~40% Voltage or Current Component | |||
U-I | U-U (Polyphase Operation) |
U-I | U-U (Polyphase Operation) |
|
16Hz ~ 69Hz | 0.003° | 0.005° | 0.010° | 0.010° |
69Hz ~ 180Hz | 0.007° | 0.007° | 0.017° | 0.017° |
180Hz ~ 450Hz | 0.018° | 0.025° | 0.050° | 0.050° |
450Hz ~ 850Hz | 0.033° | 0.043° | 0.070° | 0.070° |
850Hz ~ 3kHz | 0.115° | 0.150° | 0.200° | 0.200° |
3kHz ~ 6kHz | 0.230° | 0.300° | 0.450° | 0.450° |
Current Range | Power accuracy when current is 90% of the range (ppm) | Power accuracy when current is 50% of the range (ppm) | |||||
When voltage is 62% ~ 70% of the range | 650V & 1008V Range: 70%~75% | When voltage is 7% ~ 100% of the range | 650V&1008V Range: 70%~75% | ||||
23 V~ 90 V Range | 180 V &360 V Range | 23 V~ 90 V Range | 180 V & 360 V Range | ||||
0 ~ 2A | 62 | 64 | 64 | 72 | 74 | 74 | |
5 A~ 20A | 65 | 66 | 66 | 74 | 75 | 75 |
Current Range | Power accuracy when current is 90% of the range (ppm) | Power accuracy when current is 50% of the range (ppm) | |||||
When voltage is 62% ~ 70% of the range | 650V &1008V Range: 70%~75% | When voltage is 7% ~ 100% of the range | 650V&1008V Range: 70%~75% | ||||
23 V~ 90 V Range | 180 V& 360 V Range | 23 V~ 90 V Range | 180 V & 360 V Range | ||||
0 ~ 5A | 93 | 94 | 94 | 100 | 101 | 101 | |
10 A~ 20A | 95 | 96 | 96 | 102 | 102 | 102 |
Current Range | Power accuracy when current is 90% of the range (ppm) | Power accuracy when current is 50% of the range (ppm) | |||||
When voltage is 62% ~ 70% of the range | 650V & 1008V Range: 70%~75% | When voltage is 7% ~ 100% of the range | 650V& 1008V Range: 70%~75% | ||||
23 V~ 90 V Range | 180 V& 360 V Range | 23 V~ 90 V Range | 180 V & 360 V Range | ||||
0 ~ 5A | 97 | 98 | 98 | 103 | 105 | 105 | |
10 A~ 20A | 98 | 99 | 99 | 105 | 105 | 105 |
Maximum Frequency | 5MHz |
Maximum Frequency | 100 ns |
Maximum Count Per Channel | 232-1 |
Input Low Level Max Value | 1V |
Input High Level Max Value | 3V |
Internal Pull-Up Value | 135Ω and 940Ω to 4.5V nominal |
Approximately equivalent to 150Ω/1kΩ to 5V nominal | |
Max Input Voltage | 28V |
Min Input Voltage | 0V |
Drive | Open-collector, optional internal 470Ω pull-up |
Frequency Range | 1mHz to 5MHz |
Frequency Accuracy | ±(10ppm+100nHz) |
External Pull-Up Voltage | 30V max |
Sink Current | Max 150mA |
Drive | Open-drain |
Internal Pull-Up | The value same as gate input |
External Pull-Up Voltage | 30V max |
Sink Current | Max 1A |
Timing Accuracy | ± (10ppm+100ns) |
Gate-control Mode Accuracy | ± (10ppm+100ns) |
Packet Mode Accuracy (Ppm) | ± (Output power (ppm)+10ppm+110 000/test period) |
Set Range | ± 30% of setting value |
Flicker Modulation Depth Accuracy | 0.025% |
Setting Accuracy of Modulation Depth |
0.001% |
Modulation Envelope Shape | Square or sine wave |
Duty Factor (Square Wave) |
0.01% ~ 99.99%; accuracy = ±31 μs |
Modulation Unit | 0.5Hz ~ 40Hz |
Modulation Frequency Accuracy | <0.13% |
Number of Modulating Harmonics | 0 to all set harmonics |
Modulation Depth Setting Range | 0% ~ 100% of nominal harmonic voltage |
Modulation Accuracy (0% ~ ±30%) | ± 0.025% |
Setting Accuracy of Modulation Depth | 0.001% |
Waveform | Square wave or sine wave |
Duty Factor (Square Wave) | 0.01% ~ 99.99% |
Modulation Frequency Range | 0.008Hz~30Hz |
Sine Wave Modulation Frequency Accuracy | ± (50ppm+10μHz) |
Square Wave Modulation Frequency Accuracy | <1300ppm |
Modulation Frequency Setting Resolution | 0.001Hz |
Frequency Accuracy | ± 500ppm |
Amplitude Accuracy When 16Hz≤F<6kHz | ± 1% |
Amplitude Accuracy At >6kHz | 4% |
Max Value of Single Interharmonic | The maximum value of interharmonic below 2850Hz is 30% of the range |
Interharmonic Frequency Range | 16Hz ~ 9khz |
Min Duration | 1ms |
Max Duration | 1min |
Min Amplitude of Sag | 0% of nominal output |
Max Rapid Swell | Full scale value or 140% of nominal output, whichever is smaller |
Rise/Fall Time | 100 μs ~ 30s configurable |
Optional Repeat with Delay | 0~ (60s ±31 μs) |
Triggering Level Amplitude Accuracy | ±0.025%*level value |
Swell And Sag Level Amplitude Accuracy | ±0.25%*level value |
Trigger | The falling edge of TTL is consistent with the trigger delay end time and remains low level of 10μs ~ 31μs |