TI1100 Precision AC/DC Current Calibrator

Product Details
Customization: Available
After-sales Service: Life Time
Warranty: 12 months Excellent After-sales Service
Shipping & Policy
Shipping Cost: Contact the supplier about freight and estimated delivery time.
Payment Methods:
visa mastercard discover JCB diners club american express T/T
PIX SPEI OXXO PSE OZOW
  Support payments in USD
Secure payments: Every payment you make on Made-in-China.com is protected by the platform.
Refund policy: Claim a refund if your order doesn't ship, is missing, or arrives with product issues.
Manufacturer/Factory
Secured Trading Service
Gold Member Since 2021

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Plant Area
8000 square meters
Management System Certification
ISO9001:2015, ISO14001:2015, ISO45001:2016, Intellectual Property Management System Certificate, ISO 27001:2013
  • TI1100 Precision AC/DC Current Calibrator
  • TI1100 Precision AC/DC Current Calibrator
  • TI1100 Precision AC/DC Current Calibrator
  • TI1100 Precision AC/DC Current Calibrator
  • TI1100 Precision AC/DC Current Calibrator
  • TI1100 Precision AC/DC Current Calibrator
Find Similar Products
  • Overview
  • Company Profile
Overview

Basic Info.

Model NO.
TI1100
Customized
Non-Customized
ACV/DCV output
10 uA ~ 10.5 A
Interface
RS232; LAN; USB
Transport Package
Package for Exporting
Specification
5kg; 215 mm (W) x 252 mm (D) x 88 mm(H),
Trademark
TUNKIA
Origin
China

Packaging & Delivery

Package Size
30.00cm * 30.00cm * 10.00cm
Package Gross Weight
9.000kg

Product Description

                          TI1100 Precision AC/DC Current Calibrator TI1100 Precision AC/DC Current Calibrator

1. Summary

TI1100 is an AC/DC current source with high stability and precision, it is applied for calibrating ammeters on-site, e.g, industrial production lines, etc.
 

2. Features

  1. AC\DC current output: 10 μA ~ 10.5 A
  2. Standard version (TI1100-B) and advanced version (TI1100-C) are available
  3. Typical specifications of the standard version (TI1100-B):
          Frequency: DC, AC 10 Hz ~ 2 kHz.
         Typical short-term stability: 0.01%/min
         Optimum Accuracy: DC 0.02, AC 0.05.
     4. Typical specifications of the advanced version (TI1100-C):
         Frequency: DC, AC 10 Hz ~ 10 kHz
         Typical short-term stability: 0.005%/min
         Optimum Accuracy: DC 0.01, AC 0.02.
    5. Shortcut output keys.
    6. Programmable calibration schemes.
    7. RS232, LAN and USB communication interfaces.

3. Application


Calibration for current measuring instruments
TI1100 Precision AC/DC Current Calibrator
  1. Suitable for calibrating AC/DC ammeters or current measurement of multimeter.
 
 
  • Convenient operation
TI1100 Precision AC/DC Current Calibrator
  1. The front panel has 10 shortcut keys that can output the corresponding value on the screen with one key.
 
  • Calibration Scheme Setting
TI1100 Precision AC/DC Current Calibrator
  1. Multiple calibration schemes can be edited and stored for the calibration of multimeter with different specifications.


 

4. Specifications (TI1100-B)

4.1 DC Current Output

Range Current Output Resolution Short term stability (%/min) Accuracy
±(%*output +μA)
Compliance Voltage (V)
300 μA ±(10.0000 μA ~ 330.0000 μA) 100 pA 0.02 0.012 + 0.03 11 V
3 mA ±(0.300000 mA ~ 3.300000 mA) 1nA 0.01 0.012 + 0.24 11 V
30 mA ±(3.00000 mA ~ 33.00000 mA) 10 nA 0.01 0.012 + 2.4 9 V
300 mA ±(30.0000 mA ~ 330.0000 mA) 100 nA 0.01 0.012 + 24 7 V
1A ±(0.100000 A ~ 1.100000 A) 1μA 0.01 0.012 + 80 6 V
10 A ±(1.00000 A ~ 10.50000 A) 10 μA 0.01 0.012 + 800 4 V
  1. Output range: ± (10 μA ~ 10.5 A).
  2. 7 digits display.
  3. Adjustment fineness: 0.002% * RG.
  4. Ripple coefficient:<0.1%.
  5. Protection function: open circuit protection, overload protection, overheating protection.
 

4.2 AC Current Output

Range Current Output Frequency Resolution Accuracy
±(%*output +μA)
Compliance Voltage (rms)
300 μA 10.0000 μA ~ 330.0000 μA 10 Hz ~ 45 Hz 100 pA 0.05 + 0.2 7 V
45 Hz ~ 400 Hz 0.03 + 0.1
400 Hz ~ 2 kHz 0.04 + 0.2
3 mA 0.300000 mA ~ 3.300000 mA 10 Hz ~ 45 Hz 1nA 0.05 + 0.5 7 V
45 Hz ~ 400 Hz 0.03 + 0.3
400 Hz ~ 2 kHz 0.04 + 0.5
30 mA 3.00000 mA ~ 33.00000 mA 10 Hz ~ 45 Hz 10 nA 0.05 + 5 6 V
45 Hz ~ 400 Hz 0.03 + 3
400 Hz ~ 2 kHz 0.04 + 5
300 mA 30.0000 mA ~ 330.0000 mA 10 Hz ~ 45 Hz 100 nA 0.05 + 50 5 V
45 Hz ~ 400 Hz 0.03 + 30
400 Hz ~ 2 kHz 0.04 + 50
1A 0.100000 A ~ 1.100000 A 10 Hz ~ 45 Hz 1μA 0.05 + 150 4 V
45 Hz ~ 400 Hz 0.03 + 100
400 Hz ~ 2 kHz 0.04 + 150
10 A 1.00000 A ~ 10.50000 A 45 Hz ~ 400 Hz 10 μA 0.03 + 1000 2.5 V
400 Hz ~ 2 kHz 0.04 + 1500
  1. Output range: ± (10 μ A ~ 10.5 A).
  2. 7 digits display.
  3. Adjustment fineness: 0.002% * RG.
  4. Distortion: <0.3%.
  5. Protection function: open circuit protection, overload protection, overheating protection.
 

4.3 Frequency

Range Resolution Accuracy (± %*RD)
10.00000 Hz ≤ F ≤ 99.99999 Hz 0.01 mHz 0.01
100.0000 Hz ≤ F ≤ 999.9999 Hz 0.1 mHz 0.01
1.000000 kHz ≤ F ≤ 2.000000 kHz 1 mHz 0.01
 

5. Specifications (TI1100-C)

5.1 DC Current Output

Range Current Output Resolution 24-hour Stability (Tcal±1°C) Accuracy
1year(Tcal±5°C)
Compliance Voltage
±(ppm*output + μA)
100 μA ±(10.0000 μA ~ 110.0000 μA) 100 pA 50 + 0.01 100 + 0.02 10 V
1 mA ±(0.100000 mA ~ 1.100000 mA) 1nA 40 + 0.02 80 + 0.04 10 V
10 mA ±(1.00000 mA ~ 11.00000 mA) 10 nA 30 + 0.2 60 + 0.4 9 V
100 mA ±(10.0000 mA ~ 110.0000 mA) 100 nA 30 + 2 60 + 4 7 V
1 A ±(0.100000 A ~ 1.100000 A) 1 μA 30 + 20 60 + 40 4 V
10 A ±(1.00000 A ~ 10.50000 A) 10 μA 50 + 300 100 + 500 4 V
  1. Output range: ± (10 μA ~ 10.5 A).
  2. 7 digits display.
  3. Adjustment fineness: 0.002% * RG.
  4. Ripple coefficient: <0.1%.
  5. Protection function: open circuit protection, overload protection, overheating protection
 

5.2 AC Current Output

Range Current Output Frequency Resolution Accuracy
 ±(%*output +μA)
Compliance Voltage (rms)
100 μA 10.0000 μA ~ 110.0000 μA 10 Hz ~ 45 Hz 100 pA 800 + 0.04 6.5V
45 Hz ~ 2 kHz 600 + 0.04
2 kHz ~ 10 kHz 800 + 0.1
1 mA 0.100000 mA ~ 1.100000 mA 10 Hz ~ 45 Hz 1 nA 300 + 0.1 6.5V
45 Hz ~ 2 kHz 200 + 0.1
2 kHz ~ 10 kHz 300 + 0.2
10 mA 1.00000 mA ~ 11.00000 mA 10 Hz ~ 45 Hz 10 nA 200 + 0.8 6 V
45 Hz ~ 2 kHz 120 + 0.8
2 kHz ~ 10 kHz 300 + 2
100 mA 10.0000 mA ~ 110.0000 mA 10 Hz ~ 45 Hz 100 nA 200 + 8 5 V
45 Hz ~ 2 kHz 120 + 8
2 kHz ~ 10 kHz 300 + 20
1 A 0.100000 A ~ 1.100000 A 10 Hz ~ 45 Hz 1μA 200 + 80 3.5 V
45 Hz ~ 2 kHz 120 + 80
2 kHz ~ 10 kHz 300 + 200
10 A 1.00000 A ~ 10.50000 A 10 Hz ~ 45 Hz 10 μA 200 + 1000 2.5 V
45 Hz ~ 2 kHz 120 + 1000
2 kHz ~ 5 kHz 300 + 2000
  1. Output range: ± (10 μA ~ 10.5 A).
  2. 7 digits display.
  3. Adjustment fineness: 0.002% * RG.
  4. Distortion:<0.5%
  5. Protection function: open circuit protection, overload protection, overheating protection.
 

5.3 Frequency

Range Resolution Accuracy (± %*RD)
10.00000 Hz ≤ F ≤ 99.99999 Hz 0.01 mHz 0.01
100.0000 Hz ≤ F ≤ 999.9999 Hz 0.1 mHz 0.01
1.000000 kHz ≤ F ≤ 9.999999 kHz 1 mHz 0.01
10.00000 kHz 10 mHz 0.01
 

6. General Specifications

Power supply AC (220±22) V,(50±2) Hz
Maximum power consumption 100 VA
Warm up time Twice the time since last warmed up, to a maximum of 30 minutes.
work environment 0 ° C~55 ° C, 30% R · H~80% R · H, non-condensing
Storage environment -30 ° C~70 ° C, 10% R · H~90% R · H, non-condensing
Communication interface USB×1,LAN×1,RS232×1
Dimensions 215 mm (width) x 252 mm (depth) x 88 mm (height), excluding feets.
Display 3.5-inch color LCD
Weight About 5 kg
 
Company Profile
 

TI1100 Precision AC/DC Current Calibrator

TI1100 Precision AC/DC Current CalibratorTI1100 Precision AC/DC Current Calibrator


 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier