TI1100 Precision AC/DC Current Calibrator
1. Summary
TI1100 is an AC/DC current source with high stability and precision, it is applied for calibrating ammeters on-site, e.g, industrial production lines, etc.
2. Features
- AC\DC current output: 10 μA ~ 10.5 A
- Standard version (TI1100-B) and advanced version (TI1100-C) are available
- Typical specifications of the standard version (TI1100-B):
Frequency: DC, AC 10 Hz ~ 2 kHz.
Typical short-term stability: 0.01%/min
Optimum Accuracy: DC 0.02, AC 0.05.
4. Typical specifications of the advanced version (TI1100-C):
Frequency: DC, AC 10 Hz ~ 10 kHz
Typical short-term stability: 0.005%/min
Optimum Accuracy: DC 0.01, AC 0.02.
5. Shortcut output keys.
6. Programmable calibration schemes.
7. RS232, LAN and USB communication interfaces.
3. Application
☆ Calibration for current measuring instruments |
|
- Suitable for calibrating AC/DC ammeters or current measurement of multimeter.
|
|
|
- The front panel has 10 shortcut keys that can output the corresponding value on the screen with one key.
|
- Calibration Scheme Setting
|
- Multiple calibration schemes can be edited and stored for the calibration of multimeter with different specifications.
|
4. Specifications (TI1100-B)
4.1 DC Current Output
Range |
Current Output |
Resolution |
Short term stability (%/min) |
Accuracy ±(%*output +μA) |
Compliance Voltage (V) |
300 μA |
±(10.0000 μA ~ 330.0000 μA) |
100 pA |
0.02 |
0.012 + 0.03 |
11 V |
3 mA |
±(0.300000 mA ~ 3.300000 mA) |
1nA |
0.01 |
0.012 + 0.24 |
11 V |
30 mA |
±(3.00000 mA ~ 33.00000 mA) |
10 nA |
0.01 |
0.012 + 2.4 |
9 V |
300 mA |
±(30.0000 mA ~ 330.0000 mA) |
100 nA |
0.01 |
0.012 + 24 |
7 V |
1A |
±(0.100000 A ~ 1.100000 A) |
1μA |
0.01 |
0.012 + 80 |
6 V |
10 A |
±(1.00000 A ~ 10.50000 A) |
10 μA |
0.01 |
0.012 + 800 |
4 V |
- Output range: ± (10 μA ~ 10.5 A).
- 7 digits display.
- Adjustment fineness: 0.002% * RG.
- Ripple coefficient:<0.1%.
- Protection function: open circuit protection, overload protection, overheating protection.
4.2 AC Current Output
Range |
Current Output |
Frequency |
Resolution |
Accuracy ±(%*output +μA) |
Compliance Voltage (rms) |
300 μA |
10.0000 μA ~ 330.0000 μA |
10 Hz ~ 45 Hz |
100 pA |
0.05 + 0.2 |
7 V |
45 Hz ~ 400 Hz |
0.03 + 0.1 |
400 Hz ~ 2 kHz |
0.04 + 0.2 |
3 mA |
0.300000 mA ~ 3.300000 mA |
10 Hz ~ 45 Hz |
1nA |
0.05 + 0.5 |
7 V |
45 Hz ~ 400 Hz |
0.03 + 0.3 |
400 Hz ~ 2 kHz |
0.04 + 0.5 |
30 mA |
3.00000 mA ~ 33.00000 mA |
10 Hz ~ 45 Hz |
10 nA |
0.05 + 5 |
6 V |
45 Hz ~ 400 Hz |
0.03 + 3 |
400 Hz ~ 2 kHz |
0.04 + 5 |
300 mA |
30.0000 mA ~ 330.0000 mA |
10 Hz ~ 45 Hz |
100 nA |
0.05 + 50 |
5 V |
45 Hz ~ 400 Hz |
0.03 + 30 |
400 Hz ~ 2 kHz |
0.04 + 50 |
1A |
0.100000 A ~ 1.100000 A |
10 Hz ~ 45 Hz |
1μA |
0.05 + 150 |
4 V |
45 Hz ~ 400 Hz |
0.03 + 100 |
400 Hz ~ 2 kHz |
0.04 + 150 |
10 A |
1.00000 A ~ 10.50000 A |
45 Hz ~ 400 Hz |
10 μA |
0.03 + 1000 |
2.5 V |
400 Hz ~ 2 kHz |
0.04 + 1500 |
- Output range: ± (10 μ A ~ 10.5 A).
- 7 digits display.
- Adjustment fineness: 0.002% * RG.
- Distortion: <0.3%.
- Protection function: open circuit protection, overload protection, overheating protection.
4.3 Frequency
Range |
Resolution |
Accuracy (± %*RD) |
10.00000 Hz ≤ F ≤ 99.99999 Hz |
0.01 mHz |
0.01 |
100.0000 Hz ≤ F ≤ 999.9999 Hz |
0.1 mHz |
0.01 |
1.000000 kHz ≤ F ≤ 2.000000 kHz |
1 mHz |
0.01 |
5. Specifications (TI1100-C)
5.1 DC Current Output
Range |
Current Output |
Resolution |
24-hour Stability (Tcal±1°C) |
Accuracy 1year(Tcal±5°C) |
Compliance Voltage |
±(ppm*output + μA) |
100 μA |
±(10.0000 μA ~ 110.0000 μA) |
100 pA |
50 + 0.01 |
100 + 0.02 |
10 V |
1 mA |
±(0.100000 mA ~ 1.100000 mA) |
1nA |
40 + 0.02 |
80 + 0.04 |
10 V |
10 mA |
±(1.00000 mA ~ 11.00000 mA) |
10 nA |
30 + 0.2 |
60 + 0.4 |
9 V |
100 mA |
±(10.0000 mA ~ 110.0000 mA) |
100 nA |
30 + 2 |
60 + 4 |
7 V |
1 A |
±(0.100000 A ~ 1.100000 A) |
1 μA |
30 + 20 |
60 + 40 |
4 V |
10 A |
±(1.00000 A ~ 10.50000 A) |
10 μA |
50 + 300 |
100 + 500 |
4 V |
- Output range: ± (10 μA ~ 10.5 A).
- 7 digits display.
- Adjustment fineness: 0.002% * RG.
- Ripple coefficient: <0.1%.
- Protection function: open circuit protection, overload protection, overheating protection
5.2 AC Current Output
Range |
Current Output |
Frequency |
Resolution |
Accuracy ±(%*output +μA) |
Compliance Voltage (rms) |
100 μA |
10.0000 μA ~ 110.0000 μA |
10 Hz ~ 45 Hz |
100 pA |
800 + 0.04 |
6.5V |
45 Hz ~ 2 kHz |
600 + 0.04 |
2 kHz ~ 10 kHz |
800 + 0.1 |
1 mA |
0.100000 mA ~ 1.100000 mA |
10 Hz ~ 45 Hz |
1 nA |
300 + 0.1 |
6.5V |
45 Hz ~ 2 kHz |
200 + 0.1 |
2 kHz ~ 10 kHz |
300 + 0.2 |
10 mA |
1.00000 mA ~ 11.00000 mA |
10 Hz ~ 45 Hz |
10 nA |
200 + 0.8 |
6 V |
45 Hz ~ 2 kHz |
120 + 0.8 |
2 kHz ~ 10 kHz |
300 + 2 |
100 mA |
10.0000 mA ~ 110.0000 mA |
10 Hz ~ 45 Hz |
100 nA |
200 + 8 |
5 V |
45 Hz ~ 2 kHz |
120 + 8 |
2 kHz ~ 10 kHz |
300 + 20 |
1 A |
0.100000 A ~ 1.100000 A |
10 Hz ~ 45 Hz |
1μA |
200 + 80 |
3.5 V |
45 Hz ~ 2 kHz |
120 + 80 |
2 kHz ~ 10 kHz |
300 + 200 |
10 A |
1.00000 A ~ 10.50000 A |
10 Hz ~ 45 Hz |
10 μA |
200 + 1000 |
2.5 V |
45 Hz ~ 2 kHz |
120 + 1000 |
2 kHz ~ 5 kHz |
300 + 2000 |
- Output range: ± (10 μA ~ 10.5 A).
- 7 digits display.
- Adjustment fineness: 0.002% * RG.
- Distortion:<0.5%
- Protection function: open circuit protection, overload protection, overheating protection.
5.3 Frequency
Range |
Resolution |
Accuracy (± %*RD) |
10.00000 Hz ≤ F ≤ 99.99999 Hz |
0.01 mHz |
0.01 |
100.0000 Hz ≤ F ≤ 999.9999 Hz |
0.1 mHz |
0.01 |
1.000000 kHz ≤ F ≤ 9.999999 kHz |
1 mHz |
0.01 |
10.00000 kHz |
10 mHz |
0.01 |
6. General Specifications
Power supply |
AC (220±22) V,(50±2) Hz |
Maximum power consumption |
100 VA |
Warm up time |
Twice the time since last warmed up, to a maximum of 30 minutes. |
work environment |
0 ° C~55 ° C, 30% R · H~80% R · H, non-condensing |
Storage environment |
-30 ° C~70 ° C, 10% R · H~90% R · H, non-condensing |
Communication interface |
USB×1,LAN×1,RS232×1 |
Dimensions |
215 mm (width) x 252 mm (depth) x 88 mm (height), excluding feets. |
Display |
3.5-inch color LCD |
Weight |
About 5 kg |