Secured Trading Service

TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software

Tunkia Co., Ltd.
  • Audited Supplier
  • 6 yrs

Contact Supplier

Product Highlights

  • High Precision Measurement: Accuracy class 0.01 with stability reaching 30 ppm/min and annual error variation better than 50 ppm.
  • Automatic Calibration Software: Enables semi-automatic or fully automatic calibration, data recording, and report generation.
  • Intelligent Probe Positioning: Features stepper motors for Y/Z axis movement and rotation to determine optimal test points.
  • Magnetic Shielding Cavity: Internal field less than 10^-6 T, used to calibrate zero drift of magnetometers.
  • Dual Channel Current Source: Outputs dual DC current to excite magnetic field coil and electromagnet respectively.
  • Comprehensive Warranty: Provides 365 days warranty coverage for the equipment.
  • OEM and ODM Services: Offers customized manufacturing services with 120 R&D engineers available.

Shipping

Shipping Cost:
Contact the supplier about freight and estimated delivery time.

Online Trading Protection

Payment Guarantee
T/T PayPal ApplePay GooglePay visa mastercard american express discover JCB diners club PIX SPEI OXXO PSE OZOW Platform-protected payments with refund support for eligible order issues.
Platform Logistics
Clearer shipment tracking with platform-supported logistics.
Inspection Service
Optional pre-shipment inspection for quality and quantity checks.
After-Sales & Dispute Handling
Platform-assisted dispute resolution, including refunds or returns when applicable.

Only orders placed and paid through Made-in-China.com Secured Trading are eligible for payment protection and the corresponding platform services.

Product Images of TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software

TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software pictures & photos
TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software pictures & photos
TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software pictures & photos
TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software pictures & photos
TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software pictures & photos
TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software pictures & photos
TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
  • TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
  • TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
    TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
  • TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
    TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
  • TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
    TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
  • TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
    TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
  • TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
    TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
  • TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software
    TM9000 Tesla meter Gauss Meter Magnetometers  Automatic Calibration Equipment with software

Find Similar Products

Product Description

TM9000 Calibration Apparatus for MagnetometersTM9000 Tesla meter Gauss Meter Magnetometers Automatic Calibration Equipment with software
*The figure above is for reference, actual delivery may very slightly

1. Summary

TM9000 is a set of high precision, multi-function, intelligent magnetic parameter measurement standard device. It is composed of high-precision DC current standard source, standard electromagnet, standard Helmholtz coil, precision magnetometer, magnetic shielding cavity, automatic calibration software, etc. It is suitable for provincial and municipal metrology laboratories to set up magnetic field measurement testing standards and carry out digital or pointer type DC magnetometer calibration.
 

2. Features

  1. A standard source of bipolar excitation current equipped with dual channels. 
  2. The stability of the current source reaches 30 ppm /min.
  3. Accuracy class 0.01, the annual error variation is better than 50 ppm.
  4. Adjustment fineness is a minimum of 5ppm.
  5. Magnetic field coil generates a standard magnetic field of ± (0.1 mT to 100 mT).
  6. Electromagnets generate a highly stable magnetic field of ± (10 mT ~ 2.5 T).
  7. A custom-developed precision magnetometer of Tunkia acts as a standard meter for electromagnets.
  8. Both the magnetic field coil and the electromagnet are equipped with intelligent probe position adjustment devices.
  9. Equipped with a magnetically shielded cavity that is used to calibrate the zero drift of the magnetometer.
  10. Equipped with a portable measurement and control console for easy manual control of the magnetic field output.
  11. The system uses a modular integrated design.
  12. Comes with dedicated calibration software.

3. Instrument Appearance

TM9000 Tesla meter Gauss Meter Magnetometers Automatic Calibration Equipment with software

☆ Overall Introduction
NO. Name Description
1 Helmholtz coils The magnetic field of ±(0.1MT ~ 100mT) can be generated by DC standard current source excitation.
2 Computer and Software Realize semi-automatic or fully automatic calibration, data recording, report generation.
3 Printer Print the resulting calibration report.
4 Precision magnetometer Can be used as a standard meter to calibrate the magnetic field of an electromagnet.
5 Calibrated magnetometer The position of the DUT.
6 Electromagnet The magnetic field of ±(10mT ~ 2.5T) can be generated by DC standard current source excitation.
7 Workbench Placing equipment, proofread table, computer, printer, etc.
8 Measurement and control platform Manual control of excitation current or magnetic field output.
9 Power switch The green button is on, and the red button is off.
10 DCI standard source Produces high precision DC current for electromagnet or Helmholtz coil excitation.
11 Probe position controller Each stepper motor with 13 RS232 interfaces is connected to the probe positioning device.

4. Equipment Composition

  • High Precision DC Current Standard Source
TM9000 Tesla meter Gauss Meter Magnetometers Automatic Calibration Equipment with software
  1. High-precision bipolar current standard source, which can output dual DC current to excite the magnetic field coil and the electromagnet respectively. Accuracy is up to Class 0.01.
  2. Output stability is better than 30 ppm/min, annual error variation is better than 50 ppm.
  3. The current source can be continuously tunable by program control and has a small adjustment fineness, which facilitates the calibration of pointer magnetometer.
  4. Large power output and it can drive a magnetic field coil or electromagnet for a long time at full load.
Specifications
Dual current range 50 mA, 200 mA, 1 A, 5 A, 20 A
Output range ± 5 mA~± 24 A
Adjustment fineness 5 ppm*RG
Short term stability 30 ppm/min
Accuracy 0.006%+ 0.004%
±(ppm of reading + ppm of range) [1]
Full scale linearity < 10 pp
Annual error variation < 50 ppm
Maximum load voltage 80 V
Protection function Open circuit protection, overload protection function
Power supply AC ( 220 ± 22 )V, ( 50 ± 2 )Hz, Max. power comsumption: < 2 kVA
Note [1] : (ppm = parts per million) (e.g., 10ppm = 0.001%).
  • TM9000-HC Standard Magnetic Field with Helmholtz Coil
TM9000 Tesla meter Gauss Meter Magnetometers Automatic Calibration Equipment with software
  1. A standard magnetic field of ± (0.1 mT ~ 100 mT) can be generated by excitation with a high-precision DC standard current source.
  2. Calibration of a DC magnetometer with a radial or axial probe.
  3. The stability and accuracy of a magnetic field depends entirely on the excitation current, and can be directly regarded as a standard magnetic field.
  4. Two sets of probe position intelligent adjustment devices are installed on the coil, which drive the probe to move and rotate in the Y/Z axis by a precision stepper motor, and determine the best test point by intelligent adjustment and positioning by software.
  5. The coil is not only suitable for DC magnetometer calibration, but also suitable for various research institutes, colleges, enterprises to conduct electromagnetic interference simulation experiments, material magnetic detection experiments, etc., which is widely used.
Specifications
Field current ± ( 5 mA~20 A )
Scope ± ( 0.1 mT~100 mT ) 
Uniformity The uniform field within 20 mm is better than 300 ppm
Linearity The magnetic field is proportional to the excitation current
The temperature influence The coil constant K changes little with temperature
Inner hole diameter 200 mm
Size 680 mm × 700 mm × 450 mm (W x D x H)
Weight About 120 kg
 
  • TM9000-EM Standard Magnetic Field Electromagnet
TM9000 Tesla meter Gauss Meter Magnetometers Automatic Calibration Equipment with software
  1. The electromagnet can be excited by a high-precision DC standard current source to generate a stable magnetic field of ± (10 mT ~ 2.5T).
  2. The reference value is measured by a standard table.
  3. There is a good linear relationship between the magnetic field and the current, and the stability of the magnetic field depends on the excitation current.
  4. The TM6140B precision magnetometer (Class 0.05) is used as the standard meter.
  5. Equipped with two sets of probe position intelligent adjustment device, the probe is driven to move and rotate in the Y/Z axis by a precision stepper motor, and the optimal test point is determined by intelligent adjustment and positioning of software.
  6. The electromagnet base is equipped with rollers on each of its four legs.
  7. The electromagnet has a small input power/magnetic field ratio and can work stably for a long time without water cooling and heat dissipation.
Specifications
Field current ± ( 5 mA~24 A )
Scope ± ( 10 mT~2.5 T ) 
Uniformity The uniform field within 5 mm is better than 300 ppm
A column diameter 180 mm
Polar surface diameter 60 mm
Air gap 10 mm
Size 710 mm × 885 mm × 560 mm (W x D x H)
Weight About 500 kg
  • TM6140B Precision Magnetometer
TM9000 Tesla meter Gauss Meter Magnetometers Automatic Calibration Equipment with software
  1. Equipped with high-sensitivity, low-drift Hall sensors and applied advanced digital signal processing techniques.
  2. DC magnetic field is measured from 0 to 3000 mT, accuracy Class 0.05.
  3. As a standard meter for a magnetometer calibration device, calibrating DC magnetometer.
Range Resolution Accuracy
±(A ppm of reading + B)
The temperature coefficient
± ppm/ºC
Zero drift
± μT/h
3 mT 1 nT 0.1% + 100 μT 50 15
30 mT 10 nT 0.05% + 100 μT 50 20
300 mT 100 nT 0.05% + 100 μT 50 50
3000 mT 1 μT 0.05% + 150 μT 50 75

  • TM2800 Magnetic Shielded Cavity
 
 
TM9000 Tesla meter Gauss Meter Magnetometers Automatic Calibration Equipment with software
 
  1. The shielding frame is made of a shielding material with high magnetic conductivity, which removes the effects of the geomagnetic field.
  2. The internal magnetic field <10-6 T, which can be directly regarded as zero magnetism when calibrated.
  3. Small, lightweight (about 1kg), compact and easy to carry.
  4. The device is suitable for calibrating the zero point of a standard or calibrated magnetometer.


 
  • Magnetic Field Probe Position Controller
TM9000 Tesla meter Gauss Meter Magnetometers Automatic Calibration Equipment with software
  1. The device has 13 RS232 interfaces, 12 of which are used to control a total of 12 precision stepper motors on four probe positioning systems (two for each field coil and two for the electromagnet).
  2. The other RS232 is used for computer communication, controlling the stepper motor by software or wireless mouse remote control, driving the Hall probe to move or rotate at Y/Z axis and W Angle to determine the best magnetic field test point.
  3. The four sets of probe position intelligent adjustment devices are made of non-magnetic materials.
Axis Range of stroke/Angle Adjustment fineness
Y 50 mm 0.1 mm
Z 60 mm 0.1 mm
W ± 12° 0.1°
 
  • Special Calibration Software
TM9000 Tesla meter Gauss Meter Magnetometers Automatic Calibration Equipment with software
  1. The software automatically recognizes whether an accessory is communicating properly with the PC, and visually displays in the configuration interface.
  2. The selected excitation mechanism is visually displayed in the interface of the software and can be freely switched.
  3. The calibration parameter units: mT, G, kG, Oe, kOe, A/m, kA/m.
  4. The software controls the Y axis, Z axis, and W axis of the standard probe and the probe to adjust the best test point of the probe, or enter the wireless mouse control mode.
  5. Automatically prompts you to switch between the range and the calibrated value of the input, and automatically enters the standby state if no input is entered within 2 minutes.
  6. Automatically calculates zero drift, which exceeds the permissible zero drift and is marked in red.
  7. An error is automatically calculated and compared to the permissible error, warning you with a red light for an excess value.
  8. Test data can be generated and exported as a calibration report.
 

5. General Specifications

Power supply AC ( 220 ± 22 ) V, ( 50 ± 2 ) Hz
Temperature performance Working temperature: 0°C~50°C;
Storage temperature: -20°C~70°C
Humidity performance Working humidity: 40% ~ 80% R·H no condensation;
Storage humidity: < 80% R·H, no condensation
 
 
Company Profile

TM9000 Tesla meter Gauss Meter Magnetometers Automatic Calibration Equipment with softwareTM9000 Tesla meter Gauss Meter Magnetometers Automatic Calibration Equipment with software
TM9000 Tesla meter Gauss Meter Magnetometers Automatic Calibration Equipment with software

 

About the Supplier

This supplier has been audited by SGS. Verify the report on the official SGS website with ID "QIP-ASI2191183". Items marked "verified" are certified.

Watch Video
Watch Video
Tunkia Co., Ltd.
Tunkia Co., Ltd.
Tunkia Co., Ltd.
Tunkia Co., Ltd.
Tunkia Co., Ltd.
Tunkia Co., Ltd.
Tunkia Co., Ltd.
Tunkia Co., Ltd.

About Our Factory & Business Background

verified Business Type
Manufacturer/Factory
Address
No. 16 Panpan Road, Changsha, Hunan, China
verified Plant Area
8000 square meters
verified Number of Employees
363
verified Registered Capital
25,000,000 RMB
verified Terms of Payment
LC, T/T, D/P, PayPal, Western Union
verified International Commercial Terms(Incoterms)
FOB, CIF, CFR, EXW
verified Average Lead Time
Peak Season Lead Time: one month
Off Season Lead Time: one month
verified Nearest Port
Changsha Port

Our Production Capability & Technical Expertise

verified Main Products
Magnetic Measurement, Electrical Measurement, Gaussmeter, Multifunction Calibrator
verified Production Machines
SMT Machine, Printing Machine, Assembly & Debugging Line, Automatic Online Function Test (EPS&SST)
verified ODM Service Available
Yes
verified OEM Service Available
Yes
verified R&D Engineers
120 people
verified Own Brand
Yes

Our Industry Experience & Global Business Record

verified Year of Establishment
2006-06-16
Export Year
2021-10-11
verified Main Markets
Europe, Southeast Asia/ Mideast, East Asia(Japan/ South Korea), Domestic
verified Number of Foreign Trading Staff
2
verified Overseas Agent/Branch
No

FAQ

What is the accuracy of the TM9000?
The system features an accuracy class of 0.01 with short-term stability better than 30 ppm/min.
Does the device come with calibration software?
Yes, it includes dedicated software for automatic calibration, data recording, and report generation.
How is the probe position adjusted?
Intelligent probe position adjustment devices with precision stepper motors control movement in Y/Z axes.
What is the warranty period?
The product comes with a 365-day warranty.
Can the equipment be customized?
Yes, OEM and ODM services are available for customized requirements.
What is the lead time for orders?
The average lead time is one month for both peak and off-peak seasons.

Send your message to this supplier

Contact Information
Recipient
Message

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now