☆ Wide range current measurement |
|
☆AC and DC comprehensive measurement |
|
☆ Precise small signal voltage measurement |
|
☆ Power calculation function |
|
☆ DC ripple/AC harmonic measurement |
|
S/N | Function declaration |
1 | With DC ripple measurement function, it can measure ripple content in the range of 1 Hz~5 kHz. It can display ripple waveform in real time, and measure RMS and P-P of ripple. |
2 | The AC electric quantity measurement also has the oscilloscope function, which can display the waveform of the measured AC signal in real time. |
3 | Calculate THD/T (harmonic vs. full wave) and THD/F (harmonic vs. fundamental wave) harmonic distortion. |
4 | The amplitude (RMS), content (%) and phase of the 2nd~63rd harmonic of each phase voltage or current can be analyzed in real time. |
5 | The spectrum of each harmonic is displayed visually in the form of histogram (fundamental wave is 100%). |
☆ Data statistics and analysis function |
|
S/N | Function description |
1 | Data statistics and analysis: Calculate the maximum (Max), minimum (Min), peak to peak (P-P), average (Avg), standard deviation (S.dev), divergence (Div), etc. of the measured electricity. |
2 | Power trend chart test: during the test cycle, draw the curve of power change with time in real time. |
3 | Normal distribution histogram: it displays the distribution of the electricity to be counted during the test period. |
Range | Resolution | Accuracy (k=2),(ppm*RD + ppm*RG or V) [1] | |||
DC | 40 Hz≤F≤200 Hz | 200 Hz<F≤1 kHz | Temperature Coefficient (/°C) @(15 ~ 30)°C |
||
100mV | 10 nV | 50 + 5 μ | 120 + 80 | 200 + 100 | 3 +0.5 μV |
300 mV | 10 nV | 30 + 6 μ | 120 + 80 | 200 + 100 | 3 + 0.6 μV |
1 V | 0.1 μV | 30 + 20 | 60 + 40 | 120 + 80 | 2.5 + 1 |
3 V | 0.1 μV | 30 + 20 | 60 + 40 | 120 + 80 | 2.5 + 1 |
10 V | 1 μV | 30 + 20 | 60 + 40 | 120 + 80 | 2.5 + 0.5 |
30 V | 1 μV | 30 + 20 | 60 + 40 | 120 + 80 | 2.5 + 1 |
100 V | 10 μV | 30 + 20 | 60 + 40 | 120 + 80 | 2.5 + 1 |
300 V | 10 μV | 30 + 20 | 120 + 80 | 200 + 100 | 3.5 + 1.5 |
1000 V | 0.1 mV | 30 + 20 | 120 + 80 | 200 + 100 | 3.5 + 1.5 |
Note [1] : (ppm = parts per million) (e.g.,10ppm = 0.001%). |
Range | Resolution | Accuracy (k=2),(ppm*RD +ppm*RG or nA) | |||
DC | 40 Hz≤F≤200 Hz | 200 Hz<F≤1 kHz | Temperature Coefficient (/°C) @ (15 ~ 30)°C |
||
100 μA | 0.1 nA | 50+ 5 nA | 200 + 100 | 250 + 150 | 3 + 0.5 n |
300 μA | 0.1 nA | 50+ 15 nA | 200 + 100 | 250 + 150 | 3 + 1.5 n |
1 mA | 1 nA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
3 mA | 1 nA | 30 + 20 | 60 + 40 | 100 +50 | 2.5 + 1 |
10 mA | 10 nA | 30 + 20 | 60 + 40 | 100 + 50 | 2.5 + 1 |
30 mA | 10 nA | 30 + 20 | 60 + 40 | 100 + 50 | 2.5 + 1 |
100 mA | 0.1 μA | 30 + 20 | 60 + 40 | 100 + 50 | 2.5 + 1 |
300 mA | 0.1 μA | 30 + 20 | 60 + 40 | 100 + 50 | 2.5 + 1 |
1 A | 1 μA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
3 A | 1 μA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
10 A | 10 μA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
30 A | 10 μA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
100 A | 0.1 mA | 30 + 20 | 200 + 100 | 250 + 150 | 2.5 + 1 |
200 A [2] | 0.1 mA | 30 + 20 | 200 + 100 | 250 + 150 | 3.5 + 1.5 |
500 A [2] | 0.1 mA | 30 + 20 | 200 + 100 | 250 + 150 | 3.5 + 1.5 |
Note [2]: 200 A and 500 A range current are options, and a current transducer such as TH0540 must be selected to expand the range. |
Range | Resolution | Accuracy (k=2),±(ppm*RD +ppm*RG or μV) | |||
DC | 40 Hz≤F≤200 Hz | 200 Hz<F≤1 kHz | Temperature Coefficient (/°C) @ (15 ~ 30)°C |
||
1 mV | 1 nV | 40 + 0.2 μ | - | - | 4 + 0.02 μ |
10 mV | 10 nV | 40 +0.5 μ | 120 + 80 | 200 + 100 | 3 + 0.05 μ |
100 mV | 100 nV | 40 + 10 | 80 + 40 | 100 + 50 | 3 + 1.0 |
1 V | 1 μV | 40 + 10 | 80 + 40 | 100 + 50 | 2 + 0.5 |
10 V | 10 μV | 40 + 10 | 80 + 40 | 100 + 50 | 2 + 0.5 |
Type | Range | Accuracy (k=2) | |
40Hz ≤ F ≤ 200 Hz | 200 Hz < F ≤ 1 kHz | ||
Frequency | 40.000 00 Hz~1000.000 Hz | 20 ppm | 20 ppm |
Phase | 0.000 0°~359.999 9° | 0.02° | 0.05° |
DC Power/Electric Energy | Measuring Range | Combination of DC voltage and DC current |
Accuracy | Uncertainty of DC voltage+uncertainty of DC current (or small signal voltage) | |
AC Power/Electric Energy | Measuring Range | Combination of AC voltage and AC current |
Accuracy | Uncertainty of AC voltage+uncertainty of AC current,|PF|=1 | |
Electric Energy Pulse | Standard Electric Energy Pulse Input | directly receive the photoelectric pulse of electric energy of the tested meter |
Standard Electric Energy Pulse Output | Signal amplitude is 5V, TTL level FH=60 kHz × Power value ÷ voltage range value ÷ current range value FL=60 kHz × Power value ÷ voltage range value ÷ current range value ÷ 10000 |
|
Electric energy error display | Automatic display, resolution 0.001% |
Power Supply | AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz,Maximum power consumption:100 VA |
Temperature Performance | Working temperature:15°C~35°C; Storage temperature:-10°C~50°C |
Humidity Performance | Working humidity:< 80% @ 30°C,< 70% @ 40°C,< 40% @ 50°C Storage humidity:(20%~80%) R·H,No condensation |
Interface | RS232,RJ45 |