1 | Observation, Mark, Power on inspection | Observation items | ||
2 | Accuracy Test | Initial intrinsic error test | √ | |
3 | Starting current test | √ | ||
4 | Test of no-load | √ | ||
5 | Constant test | √ | ||
6 | Combined error of energy indication value of electronic indicator display | √ | ||
7 | Clock error test powered by power supply | √ | ||
8 | Clock error test using standby power supply | √ | ||
9 | Error consistency test | √ | ||
10 | Variation requirement test | √ | ||
11 | Load current rise and fall | √ | ||
12 | Repeat test | √ | ||
13 | Electromagnetic Test | Voltage dips and interruptions | Optional-Fast change test of load current | |
14 | Resistance to Other Influences | Harmonics in voltage and current -5th harmonic test | √ | |
15 | Harmonics in voltage and current -Square wave test | √ | ||
16 | Harmonics in voltage and current -Cusp wave test | √ | ||
17 | Sub-harmonics in the AC current circuit -Pulse train trigger waveform test | √ | ||
18 | Odd harmonics in current circuit - 90°phase trigger wave test | √ | ||
19 | DC and even harmonics - half-wave rectification waveform test | √ | ||
20 | Voltage variation test | √ | ||
21 | Frequency variation test | √ | ||
22 | Fast change test of load current test | Optional- Fast change test of load current |
||
23 | Self-heating test | √ | ||
24 | High order harmonic test | √ | ||
25 | Electrical Performance Test | Voltage line power consumption test | Optional - Voltage line power consumption test |
|
26 | Current line power consumption test | Optional - Current line power consumption test |
||
27 | IOT Electricity Meter | Harmonic power accuracy test | Optional - Harmonic energy | |
28 | Harmonic measurement accuracy test |
Range | Resolution | Short-term Stability ( % / min ) |
Accuracy ±(ppm of reading + ppm of range) [1] |
Distortion Factor ( % ) |
||
Class 0.05 | Class 0.1 | Class 0.05 | Class 0.1 | |||
220 V | 1 mV | 0.01 | 0.02 | 200 + 50 | 300 + 200 | < 0.3 |
Note [1] : (ppm = parts per million) (e.g., 10ppm = 0.001%). |
Range | Resolution | Short-term Stability ( % / min ) |
Accuracy ±(ppm of reading + ppm of range) |
Distortion Factor ( % ) |
||
Class 0.05 | Class 0.1 | Class 0.05 | Class 0.1 | |||
10 mA | 0.1 μA | 0.02 | 0.04 | 300 + 200 | 500 + 300 | < 0.5 |
20 mA | 0.1 μA | 0.02 | 0.04 | 300 + 200 | 500 + 300 | < 0.5 |
50 mA | 0.1 μA | 0.02 | 0.04 | 200 + 50 | 300 + 200 | < 0.5 |
100 mA | 1 μA | 0.01 | 0.02 | 200 + 50 | 300 + 200 | < 0.5 |
200 mA | 1 μA | 0.01 | 0.02 | 200 + 50 | 300 + 200 | < 0.5 |
500 mA | 1 μA | 0.01 | 0.02 | 200 + 50 | 300 + 200 | < 0.5 |
1 A | 10 μA | 0.01 | 0.02 | 200 + 50 | 300 + 200 | < 0.5 |
2 A | 10 μA | 0.01 | 0.02 | 200 + 50 | 300 + 200 | < 0.5 |
5 A | 10 μA | 0.01 | 0.02 | 200 + 50 | 300 + 200 | < 0.5 |
10 A | 100 μA | 0.01 | 0.02 | 200 + 50 | 300 + 200 | < 0.5 |
20 A | 100 μA | 0.01 | 0.02 | 200 + 50 | 300 + 200 | < 0.5 |
50 A | 100 μA | 0.01 | 0.02 | 200 + 50 | 300 + 200 | < 0.5 |
100 A | 1 mA | 0.01 | 0.02 | 200 + 50 | 300 + 200 | < 0.5 |
Accuracy | Class 0.05 | Class 0.1 | |
Frequency | Measuring Range | 45 Hz~100 Hz | 45 Hz~100 Hz |
Minimum Resolution | 0.000 01 Hz | 0.000 01 Hz | |
Accuracy | ±0.005%*RD | ±0.005%*RD | |
Phase (I ≥ 50mA) |
Measuring Range | 0~360° | 0~360° |
Minimum Resolution | 0.000 1° | 0.000 1° | |
Accuracy | ±0.012° | ±0.024° |
Current Range | Power Factor | Accuracy ±(% of reading) | |
Class 0.05 | Class 0.1 | ||
50 mA ≤ I ≤120 A | 0.5L~1~0.5C | 0.05%*RD | 0.1%*RD |
10 mA ≤ I < 50 mA | 1 | 0.05%*RD | 0.1%*RD |
0.5L~1~0.5C | 0.08%*RD | 0.15%*RD | |
1 mA ≤ I < 10 mA | 1 | 0.08%*RD | 0.15%*RD |
0.5L~1~0.5C | 0.15%*RD | 0.3%*RD |
Power supply | Single phase AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz | ||||
Warm up time | 30min | ||||
Temperature performance | Operating temperature: 5°C~45°C; Storage temperature: -10°C~55°C |
||||
Humidity performance | Operating humidity: < 80% @ 30°C,< 70% @ 40°C, < 40% @ 50°C; Storage humidity: (20%~80%) R·H, no condensation. |
||||
Communication interface | RS232, RS485, LAN | ||||
Dimensions of test bench (L*W *table H / total H, mm) |
Position number | 6 positions | 12 positions | 16 positions | 24 positions |
Manual wiring | 1350*750*750/1415 | -- | -- | -- | |
Crimp wiring | 1350*750*750/1415 | 1240*750*600/1782 | 1590*750*600/1782 | 2290*750*600/1782 | |
Dimensions of computer desk (L*W*H, mm) |
No standard meter | 1000*800*750 | |||
Standard meter | 1500*800*750 |
Special Customized Function Selection Table | ||
Tick | Option function | Description |
Manual wiring | The default option is Crimping wiring (with meter rack) , Manual wiring is optional. | |
Voltage line power consumption test | Recommend one position | |
Current line power consumption test | At most three position are recommended | |
Load current rise and fall | After this function is selected, all positions support; Support voltage dip and short interruption at the same time. |
|
Standard meter | TD3100 standard meter (Class 0.05/0.1) | |
Harmonic energy | TD3100-R standard meter (Class 0.05/ 0.1) |