• TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite
  • TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite
  • TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite
  • TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite
  • TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite
  • TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite

TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite

After-sales Service: Life Time
Warranty: 12 Months Excellent After-sales Service
Application: Electromagnetic Instrument, Electrical Calibration
Customization:
Shipping Cost:

Estimated freight per unit.

about shipping cost and estimated delivery time.
Payment Method: visa mastercard discover JCB diners club american express T/T
  Initial Payment Full Payment
Currency: US$
Return&refunds: You can apply for a refund up to 30 days after receipt of the products.
From payment to delivery, we protect your trading.
TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite pictures & photos
TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite
US $13,000-14,000 / Set
Min. Order: 1 Set
Gold Member Since 2021

Suppliers with verified business licenses

Secured Trading Service
Manufacturer/Factory
  • Overview
  • Summary
  • Features
  • Application
  • Characteristics
  • Specifications
  • Ordering Information
  • Company Profile
Overview

Basic Info.

Model NO.
TD1858B
Power
Electricity
Voltage Output
+(10 mV ~ 1020 V)
Current Output
±(10 μA ~ 10.2 A)
LCD Display
With LCD Display
Customized
Non-Customized
DC voltage output
+(10 mV ~ ,-(10 mV ~ 10.4 V)
DC current output
±(10 μA ~ 10.2 A)
AC voltage output
10 mV ~ 1020 V
AC current output
10 μA ~ 10.4 A
Sinusoidal Wave frequency
10 Hz ~ 20 kHz
Simulated DC resistance
1Ω ~ 220 MΩ
Transport Package
Package for Exporting
Specification
210 mm(W)* 365 mm (D) *266 mm (H), 9.5Kg
Trademark
TUNKIA
Origin
China
Production Capacity
1000

Packaging & Delivery

Package Size
28.00cm * 45.00cm * 35.00cm
Package Gross Weight
15.000kg

Product Description

TD1858B Portable Multi-function Calibrator
TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite
Summary

TD1858 is a high precision and multi-function calibrator, integrates AC\DC voltage source, AC\DC current source, DC resistance source, etc. It is portable and very suitable for on-site calibration.

Features

DC voltage output: +(10 mV ~ 1020 V),-(10 mV ~ 10.4 V)  
DC current output: ±(10 μA ~ 10.2 A)  
AC voltage output: 10 mV ~ 1020 V  
AC current output: 10 μA ~ 10.4 A  
Sinusoidal Wave frequency: 10 Hz ~ 20 kHz  
Simulated DC resistance:1 Ω ~ 220 MΩ
2 MHz Square wave frequency output  
Thermocouple outputs (optional)  
Clamp meter calibration (optional)  

Application

 

TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite
   
Characteristics

 

Wide output range
TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite
 
 Multiple output\adjustment methods
TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite
"Rotary Knob"mode,User can setting in clockwise direction or anticlockwise direction.
TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite
Use the key 10%,10%,×10,÷10 on the front panel, can realize the adjustment of the quantity value, and can also be used to perform the relevant test of linearity.
Specifications

5.1 DC Voltage Output

Range Resolution Accuracy
 ±(ppm*RD + μV) [1]
Max Burden (mA) Ripple Factor (%)
24 hours
(23±1)°C
90 days
(23±5)°C
1 year
(23±5)°C
100mV 1 μV 40 + 10 64 + 10 80 + 10 100 <0.1
1 V 10 μV 40 + 30 64 + 30 80 + 30 100 <0.1
10 V 100 μV 40 + 300 64 + 300 80 + 300 100 <0.1
100 V 1 mV 40 + 3000 64 + 3000 80 + 3000 25 <0.1
1000 V 10 mV 40 + 30000 64 + 30000 80 + 30000 12 <0.1
Note [1]: RD is the reading value, same below.
  1. Output range: +(10 mV ~ 1020 V),-(10 mV ~ 10.4 V)
  2. 6 digits display
  3. Short-circuit and overload protection

5.2 DC Current Output

Range Resolution Accuracy
±(%*RD + μA) 
Compliance Voltage  (V) Ripple Factor (%)
24 hours
(23±1)°C
90 days
(23±5)°C
1 year
(23±5)°C
100 μA 1 nA 0.01+0.03 0.016+0.03 0.02+0.03 8 <0.5
1 mA 10 nA 0.01+0.1 0.016+0.1 0.02+0.1 8 <0.2
10 mA 100 nA 0.01+ 1 0.016+ 1 0.02+ 1 8 <0.1
100 mA 1 μA 0.01+ 10 0.016+ 10 0.02+ 10 8 <0.1
1 A 10 μA 0.01+ 100 0.016+ 100 0.02+ 100 4.2 <0.1
10 A 100μA 0.015+ 2000 0.024+ 2000 0.03+ 2000 3.9 <0.1
  1. Output range: ±(10 μA ~ 10.2 A)
  2. 6 digits display
  3. Open-circuit and overload protection

5.3 AC Voltage Output

Range Resolution Frequency
(Hz)
Accuracy
 ±(%*RD + mV)
Max Burden (mA) Harmonic (%)
24 hours
(23±1)°C
90 days
(23±5)°C
1 year
(23±5)°C
100 mV 1 μV 10 ~ 2k 0.02 + 0.03 0.032 + 0.03 0.04 + 0.03 100 <0.5
2k ~ 20k 0.03 + 0.05 0.048 + 0.05 0.06 + 0.05
1 V 10 μV 10 ~ 2k 0.015 + 0.2 0.024 + 0.2 0.03 + 0.2 100 <0.5
2k ~ 20k 0.025 + 0.5 0.040 + 0.5 0.05 + 0.5
10 V 100 μV 10 ~ 2k 0.015 + 2 0.024 + 2 0.03 + 2 100 <0.5
2k ~ 20k 0.025 + 5 0.040 + 5 0.05 + 5
100 V 1 mV 40 ~ 1k 0.03 + 30 0.048 + 30 0.06 + 30 20 <0.5
1000 V 10 mV 40 ~ 1k 0.03 + 300 0.048 + 300 0.06 + 300 12 <0.5
  1. Output range: 10 mV ~ 11 V @ 10 Hz ~ 20 kHz, 10 V~1020V @ 40 Hz~1 kHz
  2. 6 digits display
  3. Short-circuit and overload protection

5.4 AC Current Output

Range Resolution Frequency
(Hz)
Accuracy
 ±(%*RD+μA)
Compliance Voltage  (V) Harmonic (%)
24 hours
(23±1)°C
90 days
(23±5)°C
1 year
(23±5)°C
100 μA 1 nA 10 ~ 2k 0.03+0.24 0.048+0.24 0.06+0.24 6 <0.5
1 mA 10 nA 10 ~ 2k 0.025+0.5 0.04+0.5 0.05+0.5 6 <0.5
10 mA 100 nA 10 ~ 2k 0.025+5 0.04+5 0.05+ 5 6 <0.5
100 mA 1 μA 10 ~ 2k 0.025+50 0.04+50 0.05+50 6 <0.5
1 A 10 μA 10 ~ 2k 0.025+600 0.04+600 0.05+600 4 <0.5
10 A 100 μA 10 ~ 2k 0.03+9000 0.048+9000 0.06+9000 3.6 <0.5
  1. Output range: 10 μA ~ 10.4 A @ 10 Hz ~ 2 kHz
  2. 6 digits display
  3. Open-circuit and overload protection

5.5 Simulated DC Resistance

Range Resolution Accuracy,±(%*RD+Ω) Allowable Current
24 hours
(23±1)°C
90 days
(23±5)°C
1 year
(23±5)°C
10 Ω 100 μΩ 0.025 + 0.05 0.03+0.05 0.05+0.05 1mA~150 mA
100 Ω 1 mΩ 0.015+0.02 0.024+0.02 0.03+0.02 1mA~80 mA
1 kΩ 10 mΩ 0.015+0.2 0.024+0.2 0.03+0.2 0.1mA~8 mA
10 kΩ 100 mΩ 0.015+2 0.024+2 0.03+ 2 10μA~800μA
100 kΩ 1 Ω 0.015+2 0.024+2 0.03+ 20 10μA~200μA
1 MΩ 10 Ω 0.015+20 0.024+20 0.03+ 200 1μA~20 μA
10 MΩ 100 Ω 0.03 + 4000 0.048 + 4000 0.06+ 4000 0.25μA~20 μA
100 MΩ 1 kΩ 0.15 + 200k 0.24 + 200k 0.3+200k 25nA~200nA
200 MΩ 1 kΩ 0.18 + 500k 0.28 + 500k 0.35+500k 15nA~100nA
  1. Output range: 10 Ω ~ 220 MΩ
  2. 6 digits display
  3. Over-current and reverse connection protection

5.6 Sinusoidal Wave Frequency Output

Range[2] Resolution Accuracy
10.0000 Hz ≤ F ≤ 99.9999 Hz 0.1 mHz ±0.01%
100.000 Hz ≤ F ≤ 999.999 Hz 1 mHz ±0.01%
1.000 00kHz ≤ F ≤ 9.999 99kHz 10 mHz ±0.01%
10.000 0 kHz ≤ F ≤ 20.000 0 kHz 0.1 Hz ±0.01%
Note [2]: The output mode is AC voltage or AC current.
 

5.7 Pulse Frequency Output

Range [3] Resolution Accuracy Shake
1.000 00 Hz ≤ F ≤ 9.999 99 Hz 10 μHz ±(20 ppm*RD + 20 μHz) <20 ns
10.000 0 Hz ≤ F ≤ 99.999 9 Hz 0.1 mHz
100.000 Hz ≤ F ≤ 999.999 Hz 1 mHz
1.000 00 kHz ≤ F ≤ 9.999 99 kHz 10 mHz
10.000 0 kHz ≤ F ≤ 99.999 9 kHz 0.1 Hz
100.000 kHz ≤ F ≤ 999.999 kHz 1 Hz
1.000 00 MHz ≤ F ≤ 2.000 00 MHz 10 Hz
Note [3]:The output type is TTL level.
 

5.8 Thermocouple Output (optional)

Type Range[4] [5]
°C
Accuracy
 (°C) @ (23±5)°C[6]
min max 90 days 1 year
J -210 -100 0.32 0.40
-100 760 0.16 0.20
760 1200 0.28 0.35
K -200 -100 0.40 0.50
-100 120 0.20 0.25
120 1370 0.36 0.45
T -200 -150 0.80 1.00
-150 400 0.16 0.20
R -50 250 1.20 1.50
250 1760 0.72 0.90
S -50 250 1.20 1.50
250 1760 0.72 0.90
B 410 1820 1.12 1.40
N -200 -100 0.64 0.80
-100 410 0.28 0.35
410 1300 0.36 0.45
E -250 -100 0.64 0.80
-100 650 0.16 0.20
650 1000 0.20 0.25
L -200 900 0.48 0.60
U -200 900 0.60 0.75
C 0 1000 0.40 0.50
1000 2310 1.04 1.30
Note [4]: temperature display resolution: 0.01 ° C;
Note [5]: Internal resistance of output source: 10 Ω;
Note [6]: thermocouple error is not included.
Note [7]: Use external compensation, J, K, T, R, S, B, N, E conform to ITS-90 international temperature standard, L, U conform to DIN 43710-1985 standard, C conform to ASTM standard.
 
Ordering Information

 

TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite

 

Company Profile
 

TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite

TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator OnsiteTUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite

 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now

You Might Also Like

From payment to delivery, we protect your trading.
TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite pictures & photos
TUNKIA Portable 4.5 digit Digital Multimeter Multi Product Calibrator Onsite
US $13,000-14,000 / Set
Min. Order: 1 Set
Gold Member Since 2021

Suppliers with verified business licenses

Secured Trading Service
Manufacturer/Factory
Plant Area
8000 square meters
Management System Certification
ISO9001:2015, ISO14001:2015, ISO45001:2016, Intellectual Property Management System Certificate, ISO 27001:2013