TD7300 Defibrillation Pulse Energy Standard
1. Summary
TD7300 is a device applied for energy detection of defibrillator analyzer. It is used for medical equipment manufacturers and medical equipment testing institutions at each level carry out registration inspection, metrological traceability, etc.
2. Features
- Energy output:1 J~400 J
- Maximum voltage output:3700 V;
- Discharge mode: monophasic wave (MDS, MTE), biphasic wave (BTE);
- Discharge pulse duration:about 10 ms~16 ms;
- LCD touch screen, with physical keys;
- Interface: RS232
3. Application
- Manufacturer of defibrillator analyzer
- Measurement and calibration units at all levels
☆ Calibrate the Energy Specifications of Defibrillator Analyzer |
|
- Energy output:1 J~400 J, Maximum voltage output:3700 V
- Internal energy storage capacitor and equipped with internal discharge circuit, which can discharge manually or automatically after 1min.
|
4. Characteristics
☆ Convenient Operation |
|
- LCD screen, supporting touch operation.
- Digital control button: it can realize fixed-point output mode.
|
5. Specifications
Standard Energy Output |
1 J~400 J |
Maximum Voltage Output |
3700 V |
Accuracy |
±0.3%*reading |
Discharge Mode |
Monophasic Wave (MDS, MTE), Biphasic Wave (BTE) |
Discharge Pulse Duration |
10 ms~16 ms |
Charging Time |
400J Energy Charging Time ≤ 15s |
6. General Specifications
Power Supply |
AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz |
Temperature Performance |
Working temperature:0°C~45°C;Storage temperature:-20°C~70°C |
Humidity Performance |
Working humidity:< 80% @ 30°C,< 70% @ 40°C,< 40% @ 50°C Storage humidity:(20%~80%) R·H,non-condensing |
Interface |
RS232 |