After-sales Service: 24 Hours
Warranty: 365 Days
DC Current Standard Source: ± 1 mA~± 50 A
Customization:
Shipping Cost:

Estimated freight per unit.

about shipping cost and estimated delivery time.
Payment Method: visa mastercard discover JCB diners club american express T/T
  Initial Payment Full Payment
Currency: US$
Return&refunds: You can apply for a refund up to 30 days after receipt of the products.
From payment to delivery, we protect your trading.
TUNKIA TM9100 Automated Precision Magnetometer Calibration Device for PT2025  NMR Tesla meter Gauss Meter pictures & photos
TUNKIA TM9100 Automated Precision Magnetometer Calibration Device for PT2025  NMR Tesla meter Gauss Meter
US $1,000-180,000 / Piece
Min. Order: 1 Piece
Gold Member Since 2021

Suppliers with verified business licenses

Secured Trading Service
Manufacturer/Factory
  • Overview
  • Company Profile
Overview

Basic Info.

Model NO.
TUNKIA
Short Term Stability
20 ppm/min
Accuracy
0.01%
Helmholtz Coil magnetic field
0.1 mT ~ 100 mT
Electromagnet magnetic field
20 mT~2.6 T
Transport Package
Air Package
Specification
500 Kg
Trademark
TUNKIA
Origin
China
HS Code
9030899090
Production Capacity
1000 Set/Year

Packaging & Delivery

Package Size
100.00cm * 100.00cm * 100.00cm
Package Gross Weight
200.000kg

Product Description


TM9100 Precision Magnetometer Calibration Device

*The figure above is for reference, actual delivery may very slightly

1. Summary

TM9100 is a set of high-precision, multi-functional and intelligent magnetic parameter measurement standard device. It consists of high-precision DC current standard source, standard electromagnet, standard Helmholtz coil, precision magnetometer, magnetic shielding cavity, automatic calibration software, etc. It is suitable for metrology laboratories to establish precision magnetic field measurement and inspection standards and carry out calibration work for Tesla meters or PT2025 nuclear magnetic resonance magnetometers of class 0.05 and below. Reference standard: JJF 1832-2020 "(1 mT ~ 2.5 T) Magnetometer Calibration Specifications" (TUNKIA participated in the drafting).
 

2. Features

  1. Equipped with dual-channel bipolar excitation current standard source.
  2. The stability of the current source reaches 20 ppm/min.
  3. The accuracy of the current source reaches class 0.01, and the annual error change is better than 50 ppm.
  4. The magnetic field coil can generate a standard magnetic field of ± (0.1 mT~100 mT).
  5. The electromagnet can generate a highly stable magnetic field of ± (20 mT~2.6 T).
  6. Optional PT2025 or PT2026 NMR magnetometer as a standard meter.
  7. Both the magnetic field coil and the electromagnet are equipped with intelligent probe position adjustment devices.
  8. Supports fully automatic switching of multi-probe ranges of PT2026. (optional accessories)
  9. Equipped with a magnetic shielding cavity for calibrating the zero drift of the magnetometer.
  10. Equipped with a portable measurement and control console to facilitate manual control of magnetic field output.
  11. Equipped with specialized calibration software

3. Application

 Calibrate Precision Magnetometer
TUNKIA TM9100 Automated Precision Magnetometer Calibration Device for PT2025 NMR Tesla meter Gauss Meter
  1. The PT2025 nuclear magnetic resonance magnetometer(optional) is used as a standard meter, which is suitable for calibrating precision magnetometers of class 0.05 and below. 


 
Calibrate NMR Magnetometer
TUNKIA TM9100 Automated Precision Magnetometer Calibration Device for PT2025 NMR Tesla meter Gauss Meter
  1. The PT2026 nuclear magnetic resonance magnetometer is used as a standard meter, which is suitable for calibrating the PT2025 NMR magnetometer.
  2. Also suitable for calibrating precision magnetometers of class 0.05 and below.
 

4. Equipment Composition

TUNKIA TM9100 Automated Precision Magnetometer Calibration Device for PT2025 NMR Tesla meter Gauss Meter
  • High Precision DC Current Standard Source
  1. This current source is a high-precision bipolar current standard source that can output dual-channel DC current to excite the magnetic field coil and electromagnet respectively. Each current output has high accuracy and stability, and the measurement uncertainty reaches class 0.01.
  2. The current output stability is better than 20 ppm/min, and the annual error change is better than 50 ppm.
  3. The current source can be continuously adjusted through program control, and has a small adjustment fineness, which facilitates the calibration of the pointer magnetometer.
  4. It has a large power output and can drive magnetic field coils or electromagnets to work for a long time under full load.
Specifications
Dual Current Range 50 mA,200 mA,1 A,5 A,20 A,50 A
Output Range ± 1 mA~± 50 A
Adjustment Fineness 0.001%*RG
Short Term Stability 20 ppm/min
Optimal Measurement Uncertainty 0.006%*RD[1]+ 0.004%*RG[2]
Full Scale Linearity < 10 ppm
Annual Error Variation < 50 ppm
Maximum Load Voltage 180V
Protection Function Open circuit protection, overload protection function
Power Supply AC ( 380 ± 38 ) V,( 50 ± 2 ) Hz,
Max. power consumption: < 10 kVA
Note: [1] RD is the reading value, [2] RG is the range value.

TUNKIA TM9100 Automated Precision Magnetometer Calibration Device for PT2025 NMR Tesla meter Gauss Meter 
  • TM9100-HC Standard Magnetic Field with Helmholtz Coil
  1. A standard magnetic field of ± (0.1 mT ~ 100 mT) can be generated by excitation with a high-precision DC standard current source.
  2. Calibration of a DC magnetometer with a radial or axial probe.
  3. The stability and accuracy of a magnetic field depends entirely on the excitation current, and can be directly regarded as a standard magnetic field.
  4. Two sets of probe position intelligent adjustment devices are installed on the coil, which drive the probe to move and rotate in the Y/Z axis by a precision stepper motor, and determine the best test point by intelligent adjustment and positioning of software.
  5. The coil is not only suitable for DC magnetometer calibration, but also suitable for various research institutes, colleges, enterprises to conduct electromagnetic interference simulation experiments, material magnetic detection experiments, etc., which is widely used.
Specifications
Excitation Current ± ( 5 mA~20 A )
Magnetic Field Range ± ( 0.1 mT~100 mT ) 
Magnetic Field Uniformity The uniform field within 20 mm is better than 300 ppm
Magnetic Field Linearity The magnetic field is proportional to the excitation current
Temperature Influence The coil constant K changes little with temperature
Inner Hole Diameter 200 mm
Size 680 mm × 700 mm × 450 mm (W x D x H)
Weight Approx 120 kg

TUNKIA TM9100 Automated Precision Magnetometer Calibration Device for PT2025 NMR Tesla meter Gauss Meter
 
  • TM9100-EM Electromagnet for Standard Magnetic Field
  1. Two specifications of polar are optional, which can generate a stable magnetic field of up to 2.6 T.
  2. Optional PT2025 or PT2026 nuclear magnetic resonance magnetometer as a standard meter.
  3. There is a good linear relationship between magnetic field and current, and the stability of the magnetic field depends on the excitation current.
  4. Equipped with an intelligent probe position adjustment device, which can intelligently adjust the positioning through software to determine the best test point.
  5. The input power/magnetic field ratio of the electromagnet is small, and it can work stably for a long time without water cooling.
Specifications
Excitation Current ± ( 5 mA~50 A )
Polar 280 mm-140 mm 280 mm-200 mm
Magnetic Field Range ± ( 20 mT~2.6 T )  ± ( 20 mT~2.5 T )
Magnetic Field Uniformity Uniformity within 5 mm is better than 150 ppm 2.1T: Uniformity within ±10 mm is better than 250 ppm
2.1T: Uniformity within ±20 mm is better than 1000 ppm
2.5T: Uniformity within ±5 mm is better than 250 ppm
2.5T: Uniformity within ±10 mm is better than 1000 ppm
Air Gap 15 mm 16 mm
Size 700 mm × 1100 mm × 1000 mm (W x D x H)
Weight Approx 2000 kg
 
  • TM2800 Magnetic Shielded Cavity
 
 
TUNKIA TM9100 Automated Precision Magnetometer Calibration Device for PT2025 NMR Tesla meter Gauss Meter
 
  1. The shielding frame is made of a shielding material with high magnetic conductivity, which removes the effects of the geomagnetic field.
  2. The internal magnetic field <10-6 T, which can be directly regarded as zero magnetism when calibrated.
  3. Small, lightweight (about 1kg), compact and easy to carry.
  4. The device is suitable for calibrating the zero point of a standard or calibrated magnetometer.


 
  • Magnetic Field Probe Position Controller
TUNKIA TM9100 Automated Precision Magnetometer Calibration Device for PT2025 NMR Tesla meter Gauss Meter
  1. The software automatically recognizes whether an accessory is communicating properly with the PC, and visually displays in the configuration interface.
  2. The selected excitation mechanism is visually displayed in the interface of the software and can be freely switched.
  3. The calibration parameter units: mT, G, kG, Oe, kOe, A/m, kA/m.
  4. The software controls the Y axis, Z axis, and W axis of the standard probe and the probe to adjust the best test point of the probe, or enter the wireless mouse control mode.
  5. Automatically prompts you to switch between the range and the calibrated value of the input, and automatically enters the standby state if no input is entered within 2 minutes.
  6. Automatically calculates zero drift, which exceeds the permissible zero drift and is marked in red.
  7. An error is automatically calculated and compared to the permissible error, warning you with a red light for an excess value.
  8. Test data can be generated and exported as a calibration report.
Axis Range of stroke/Angle Adjustment fineness
Y 50 mm 0.1 mm
Z 60 mm 0.1 mm
W ± 12° 0.1°
 
  • Special Calibration Software
TUNKIA TM9100 Automated Precision Magnetometer Calibration Device for PT2025 NMR Tesla meter Gauss Meter
  1. The software automatically recognizes whether an accessory is communicating properly with the PC, and visually displays in the configuration interface.
  2. The selected excitation mechanism is visually displayed in the interface of the software and can be freely switched.
  3. The calibration parameter units: mT, G, kG, Oe, kOe, A/m, kA/m.
  4. The software controls the Y axis, Z axis, and W axis of the standard probe and the probe to adjust the best test point of the probe, or enter the wireless mouse control mode.
  5. Automatically prompts you to switch between the range and the calibrated value of the input, and automatically enters the standby state if no input is entered within 2 minutes.
  6. Automatically calculates zero drift, which exceeds the permissible zero drift and is marked in red.
  7. An error is automatically calculated and compared to the permissible error, warning you with a red light for an excess value.
  8. Test data can be generated and exported as a calibration report.
 

5. General Specifications

Power supply AC ( 220 ± 22 ) V, ( 50 ± 2 ) Hz
Temperature performance Working temperature: 0°C~50°C;
Storage temperature: -20°C~70°C
Humidity performance Working humidity: 40% ~ 80% R·H no condensation;
Storage humidity: < 80% R·H, no condensation
 
 
Company Profile

TUNKIA TM9100 Automated Precision Magnetometer Calibration Device for PT2025 NMR Tesla meter Gauss MeterTUNKIA TM9100 Automated Precision Magnetometer Calibration Device for PT2025 NMR Tesla meter Gauss Meter
TUNKIA TM9100 Automated Precision Magnetometer Calibration Device for PT2025 NMR Tesla meter Gauss Meter

 

 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now

You Might Also Like

From payment to delivery, we protect your trading.
TUNKIA TM9100 Automated Precision Magnetometer Calibration Device for PT2025  NMR Tesla meter Gauss Meter pictures & photos
TUNKIA TM9100 Automated Precision Magnetometer Calibration Device for PT2025  NMR Tesla meter Gauss Meter
US $1,000-180,000 / Piece
Min. Order: 1 Piece
Gold Member Since 2021

Suppliers with verified business licenses

Secured Trading Service
Manufacturer/Factory
Plant Area
8000 square meters
Management System Certification
ISO9001:2015, ISO14001:2015, ISO45001:2016, Intellectual Property Management System Certificate, ISO 27001:2013