Customization: | Available |
---|---|
After-sales Service: | Life Time |
Warranty: | 12 Months Excellent After-sales Service |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
☆ Calibrate AC\DC electrical measuring instruments |
DCV output: +(10 mV ~ 1020 V), -(10 mV ~ 10.4 V)
DCI output: ±(10 μA ~ 10.2 A)
ACV output: 10 mV ~ 1020V, 10 Hz ~ 20 kHz
ACI output: 10 μA ~ 10.4 A, 10 Hz ~ 2 kHz
RES output: 1 Ω ~ 220 MΩ
Frequency output:1 Hz ~ 2 MHz
Calibrate digital multimeter, AC/DC voltmeter, AC/DC ammeter, DC ohmmeter, frequency meter.
|
☆ Calibrate clamp ammeter (optional) |
Equivalent to 1000AT current through input 20A current to the 50T Coil, which is suitable for calibrating DC clamp meters.
|
☆ Calibrate thermocouple (optional) |
Type of Thermocouple (TC) Output:J,K,T,R,S,B,N,E,L,U,C.
|
☆ Front\Side Panel | |
Item | Description |
1 | LCD touch screen. |
2 | AC/DC voltage and Simulated DC resistance output terminals. |
3 | AC/DC current output terminals. |
4 | Pulse frequency output terminals. |
5 | Thermocouple output/measurement terminals. |
6 | Operation panel. |
7 | Foldable feets. |
☆ Rear Panel | |
Item | Description |
1 | Embedded handle. |
2 | USB,LAN,RS232 interface. |
3 | Cooling vents. |
4 | Power switch. |
5 | AC 220V power interface. |
6 | Chassis ground binding post. |
☆ Wide output range |
☆ Excellent portability |
The overall weight is less than 10 kg.
|
☆ Multiple output\adjustment methods |
"Rotary Knob"mode,User can setting in clockwise direction or anticlockwise direction.
|
Use the key ↑10%,↓10%,×10,÷10 on the front panel, can realize the adjustment of the quantity value, and can also be used to perform the relevant test of linearity.
|
Range | Resolution | Accuracy ±(ppm*RD + μV) [1] |
Max Burden (mA) | Ripple Factor (%) | ||
24 hours (23±1)°C |
90 days (23±5)°C |
1 year (23±5)°C |
||||
100mV | 1 μV | 40 + 10 | 64 + 10 | 80 + 10 | 100 | <0.1 |
1 V | 10 μV | 40 + 30 | 64 + 30 | 80 + 30 | 100 | <0.1 |
10 V | 100 μV | 40 + 300 | 64 + 300 | 80 + 300 | 100 | <0.1 |
100 V | 1 mV | 40 + 3000 | 64 + 3000 | 80 + 3000 | 25 | <0.1 |
1000 V | 10 mV | 40 + 30000 | 64 + 30000 | 80 + 30000 | 12 | <0.1 |
Note [1]: RD is the reading value, same below. |
Range | Resolution | Accuracy ±(%*RD + μA) |
Compliance Voltage (V) | Ripple Factor (%) | ||
24 hours (23±1)°C |
90 days (23±5)°C |
1 year (23±5)°C |
||||
100 μA | 1 nA | 0.01+0.03 | 0.016+0.03 | 0.02+0.03 | 8 | <0.5 |
1 mA | 10 nA | 0.01+0.1 | 0.016+0.1 | 0.02+0.1 | 8 | <0.2 |
10 mA | 100 nA | 0.01+ 1 | 0.016+ 1 | 0.02+ 1 | 8 | <0.1 |
100 mA | 1 μA | 0.01+ 10 | 0.016+ 10 | 0.02+ 10 | 8 | <0.1 |
1 A | 10 μA | 0.01+ 100 | 0.016+ 100 | 0.02+ 100 | 4.2 | <0.1 |
10 A | 100μA | 0.015+ 2000 | 0.024+ 2000 | 0.03+ 2000 | 3.9 | <0.1 |
Range | Resolution | Frequency (Hz) |
Accuracy ±(%*RD + mV) |
Max Burden (mA) | Harmonic (%) | ||
24 hours (23±1)°C |
90 days (23±5)°C |
1 year (23±5)°C |
|||||
100 mV | 1 μV | 10 ~ 2k | 0.02 + 0.03 | 0.032 + 0.03 | 0.04 + 0.03 | 100 | <0.5 |
2k ~ 20k | 0.03 + 0.05 | 0.048 + 0.05 | 0.06 + 0.05 | ||||
1 V | 10 μV | 10 ~ 2k | 0.015 + 0.2 | 0.024 + 0.2 | 0.03 + 0.2 | 100 | <0.5 |
2k ~ 20k | 0.025 + 0.5 | 0.040 + 0.5 | 0.05 + 0.5 | ||||
10 V | 100 μV | 10 ~ 2k | 0.015 + 2 | 0.024 + 2 | 0.03 + 2 | 100 | <0.5 |
2k ~ 20k | 0.025 + 5 | 0.040 + 5 | 0.05 + 5 | ||||
100 V | 1 mV | 40 ~ 1k | 0.03 + 30 | 0.048 + 30 | 0.06 + 30 | 20 | <0.5 |
1000 V | 10 mV | 40 ~ 1k | 0.03 + 300 | 0.048 + 300 | 0.06 + 300 | 12 | <0.5 |
Range | Resolution | Frequency (Hz) |
Accuracy ±(%*RD+μA) |
Compliance Voltage (V) | Harmonic (%) | ||
24 hours (23±1)°C |
90 days (23±5)°C |
1 year (23±5)°C |
|||||
100 μA | 1 nA | 10 ~ 2k | 0.03+0.24 | 0.048+0.24 | 0.06+0.24 | 6 | <0.5 |
1 mA | 10 nA | 10 ~ 2k | 0.025+0.5 | 0.04+0.5 | 0.05+0.5 | 6 | <0.5 |
10 mA | 100 nA | 10 ~ 2k | 0.025+5 | 0.04+5 | 0.05+ 5 | 6 | <0.5 |
100 mA | 1 μA | 10 ~ 2k | 0.025+50 | 0.04+50 | 0.05+50 | 6 | <0.5 |
1 A | 10 μA | 10 ~ 2k | 0.025+600 | 0.04+600 | 0.05+600 | 4 | <0.5 |
10 A | 100 μA | 10 ~ 2k | 0.03+9000 | 0.048+9000 | 0.06+9000 | 3.6 | <0.5 |
Range | Resolution | Accuracy,±(%*RD+Ω) | Allowable Current | ||
24 hours (23±1)°C |
90 days (23±5)°C |
1 year (23±5)°C |
|||
10 Ω | 100 μΩ | 0.025 + 0.05 | 0.03+0.05 | 0.05+0.05 | 1mA~150 mA |
100 Ω | 1 mΩ | 0.015+0.02 | 0.024+0.02 | 0.03+0.02 | 1mA~80 mA |
1 kΩ | 10 mΩ | 0.015+0.2 | 0.024+0.2 | 0.03+0.2 | 0.1mA~8 mA |
10 kΩ | 100 mΩ | 0.015+2 | 0.024+2 | 0.03+ 2 | 10μA~800μA |
100 kΩ | 1 Ω | 0.015+2 | 0.024+2 | 0.03+ 20 | 10μA~200μA |
1 MΩ | 10 Ω | 0.015+20 | 0.024+20 | 0.03+ 200 | 1μA~20 μA |
10 MΩ | 100 Ω | 0.03 + 4000 | 0.048 + 4000 | 0.06+ 4000 | 0.25μA~20 μA |
100 MΩ | 1 kΩ | 0.15 + 200k | 0.24 + 200k | 0.3+200k | 25nA~200nA |
200 MΩ | 1 kΩ | 0.18 + 500k | 0.28 + 500k | 0.35+500k | 15nA~100nA |
Range[2] | Resolution | Accuracy |
10.0000 Hz ≤ F ≤ 99.9999 Hz | 0.1 mHz | ±0.01% |
100.000 Hz ≤ F ≤ 999.999 Hz | 1 mHz | ±0.01% |
1.000 00kHz ≤ F ≤ 9.999 99kHz | 10 mHz | ±0.01% |
10.000 0 kHz ≤ F ≤ 20.000 0 kHz | 0.1 Hz | ±0.01% |
Note [2]: The output mode is AC voltage or AC current. |
Range [3] | Resolution | Accuracy | Shake |
1.000 00 Hz ≤ F ≤ 9.999 99 Hz | 10 μHz | ±(20 ppm*RD + 20 μHz) | <20 ns |
10.000 0 Hz ≤ F ≤ 99.999 9 Hz | 0.1 mHz | ||
100.000 Hz ≤ F ≤ 999.999 Hz | 1 mHz | ||
1.000 00 kHz ≤ F ≤ 9.999 99 kHz | 10 mHz | ||
10.000 0 kHz ≤ F ≤ 99.999 9 kHz | 0.1 Hz | ||
100.000 kHz ≤ F ≤ 999.999 kHz | 1 Hz | ||
1.000 00 MHz ≤ F ≤ 2.000 00 MHz | 10 Hz | ||
Note [3]:The output type is TTL level. |
Type | Range[4] [5] °C |
Accuracy (°C) @ (23±5)°C[6] |
||
min | max | 90 days | 1 year | |
J | -210 | -100 | 0.32 | 0.40 |
-100 | 760 | 0.16 | 0.20 | |
760 | 1200 | 0.28 | 0.35 | |
K | -200 | -100 | 0.40 | 0.50 |
-100 | 120 | 0.20 | 0.25 | |
120 | 1370 | 0.36 | 0.45 | |
T | -200 | -150 | 0.80 | 1.00 |
-150 | 400 | 0.16 | 0.20 | |
R | -50 | 250 | 1.20 | 1.50 |
250 | 1760 | 0.72 | 0.90 | |
S | -50 | 250 | 1.20 | 1.50 |
250 | 1760 | 0.72 | 0.90 | |
B | 410 | 1820 | 1.12 | 1.40 |
N | -200 | -100 | 0.64 | 0.80 |
-100 | 410 | 0.28 | 0.35 | |
410 | 1300 | 0.36 | 0.45 | |
E | -250 | -100 | 0.64 | 0.80 |
-100 | 650 | 0.16 | 0.20 | |
650 | 1000 | 0.20 | 0.25 | |
L | -200 | 900 | 0.48 | 0.60 |
U | -200 | 900 | 0.60 | 0.75 |
C | 0 | 1000 | 0.40 | 0.50 |
1000 | 2310 | 1.04 | 1.30 | |
Note [4]: temperature display resolution: 0.01 ° C; Note [5]: Internal resistance of output source: 10 Ω; Note [6]: thermocouple error is not included. Note [7]: Use external compensation, J, K, T, R, S, B, N, E conform to ITS-90 international temperature standard, L, U conform to DIN 43710-1985 standard, C conform to ASTM standard. |
Power supply | AC (220±22) V,(50±2) Hz |
Maximum power consumption | 500 VA |
Warm up time | Twice the time since last warmed up, to a maximum of 30 minutes. |
Temperature performance | Operating temperature:0°C~40°C Calibration temperature:18°C~28°C Storage temperature:-20°C~70°C |
Humidity performance | Operating humidity:<80% @ 30°C,<70% @ 40°C Storage humidity:(20%~80%) R·H,No condensation |
Communication interface | RS232×1,LAN×1,USB×1 |
Dimensions | 210 mm(W)× 365 mm (D) × 266 mm (H),excluding handles and feets. |
Weight | About 9.5 kg |