Customization: | Available |
---|---|
After-sales Service: | Life Time |
Warranty: | 12 Months Excellent After-sales Service |
Shipping Cost: | Contact the supplier about freight and estimated delivery time. |
---|
Payment Methods: |
|
---|---|
Support payments in USD |
Secure payments: | Every payment you make on Made-in-China.com is protected by the platform. |
---|
Refund policy: | Claim a refund if your order doesn't ship, is missing, or arrives with product issues. |
---|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
☆ Multiple output methods |
The instrument has a "fixed point output" mode, through the digital button of the operation table or click the touch screen, directly set the required output value, the instrument will automatically switch to the best range output.
|
☆ Multiple output methods |
|
When calibrating a resistance tester, it is usually necessary to select the calibration point according to the proportion of each range of the meter to be checked.
The user can easily select the calibration point of the inspected meter through the "Percentage Output Button" on the console of this instrument or the "Percentage Check Point" on the touch screen.
|
The operating area is equipped with a "rotary encoder" that increases or decreases the output by rotating it clockwise or counterclockwise.
|
|
||
S/N | Function | |
1 | Integration period: The integration period can be set in the range of 0.5 s ~ 20 s (default 1 s), which is convenient for users to set sampling data according to actual needs; | |
2 | Statistical analysis: real-time recording and statistical test data, including current average Avg, mean square deviation S.dev, peak-to-peak P-P, and maximum value Max, minimum min, test number N, time time, etc., easy for users to read intuitively; |
☆ Rich communication interface and special test software |
RS232, USB, LAN interfaces.
|
Range | Fineness Adjustment | Measurement Uncertainty (k=2, Ppm*Rd+Μω)[1] | Input Current[2] | ||
24h ( 23 ± 1 )°C |
90days ( 23 ± 5 )°C |
1year ( 23 ± 5 )°C |
|||
1 mΩ | 5 nΩ | 40 + 0.5 | 60 + 0.5 | 80 + 0.5 | 1 A~10 A |
10 mΩ | 50 nΩ | 40 + 1 | 60 + 1 | 80 + 1 | 0.5 A~5 A |
100 mΩ | 500 nΩ | 25 + 3 | 40 + 3 | 50 + 3 | 0.3A ~5 A |
1 Ω | 5 μΩ | 25 + 10 | 40 + 10 | 50 + 10 | 0.05 A~5 A |
10 Ω | 50 μΩ | 25 + 100 | 40 + 100 | 50 + 100 | 10 mA~300 mA |
100 Ω | 500 μΩ | 25 + 1000 | 40 + 1000 | 50 + 1000 | 1 mA~30 mA |
Note [1]: RD is the reading value, the same below; [2] When the current and resistance exceed the range, the resistance value will bring additional error |
Range | Fineness Adjustment | Measurement Uncertainty ( k=2, ppm*RD+Ω ) | Input Current | ||
24h (23±1)°C |
90days (23±5)°C |
1year (23±5)°C |
|||
100 Ω | 500 μΩ | 12 + 0.002 | 20 + 0.002 | 25 + 0.002 | 1 mA~80 mA |
300 Ω | 1.5 mΩ | 12 + 0.003 | 20 + 0.003 | 25 + 0.003 | 1 mA~60 mA |
1 kΩ | 5 mΩ | 12 + 0.01 | 18 + 0.01 | 20 + 0.01 | 1 mA~12 mA |
3 kΩ | 15 mΩ | 12 + 0.03 | 18 + 0.03 | 20 + 0.03 | 0.1 mA~4 mA |
10 kΩ | 50 mΩ | 12 + 0.1 | 18 + 0.1 | 20 + 0.1 | 0.1 mA~2 mA |
30 kΩ | 150 mΩ | 12 + 0.3 | 18 + 0.3 | 20 + 0.3 | 10 μA~400 μA |
100 kΩ | 500 mΩ | 15 + 1 | 24 + 1 | 30 + 1 | 10 μA~200 μA |
300 kΩ | 1.5 Ω | 17 + 3 | 28 + 3 | 35 + 3 | 1 μA~40 μA |
1 MΩ | 5 Ω | 17 + 10 | 28 + 10 | 35 + 10 | 1 μA~20 μA |
3 MΩ | 15 Ω | 30 + 30 | 48 + 30 | 60 + 30 | 0.25 μA~4 μA |
10 MΩ | 50Ω | 65 + 100 | 110 + 100 | 130 + 100 | 0.25 μA~2 μA |
Power Supply | AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz; |
Warm-up Time | Not less than 1 hour, and the warm-up time again after shutting down in the middle should not be less than 2 times the shutdown time; |
Maximum Power Consumption | 150 VA |
Temperature Performance | Working temperature: 18 °C ~ 28 °C; Storage temperature: -10 °C ~ 50 °C; |
Humidity Performance | Working humidity: (20% ~ 80%) R·H, non-condensing; Storage humidity: < 85% R·H, no condensation; |
Altitude | < 3000 m |
Weight | About 10 kg |
Communication Interface | RS 232,USB,IP |
Dimension | 400 mm ( W ) × 265 mm ( D ) × 190 mm ( H )(without feet and handles) |