| Current Sensor Test | |
| Metrology Performance | Basic accuracy error |
| Zero point output error | |
| Full range output error | |
| Linearity error | |
| Return difference | |
| repeatability error | |
| Influence Quantity Performance | Power consumption measurement |
| Zero drift | |
| Thermal zero drift (needs to be equipped with a temperature control box) | |
| Thermal sensitivity drift (needs to be equipped with a temperature control box) | |
| Overload capacity | |
| Power supply voltage impact test | |
| Load change rate (needs to be equipped with a load box) | |
| Other Performance | Measurement bandwidth (requires high-frequency constant current source) |
| Response time (requires pulse current source) | |
| ☆ Metrology Performance Test |
|
| ☆ Power Supply Impact Test And Power Consumption Measurement |
|
| ☆ Sensor Bandwidth Test (Optional) |
|
| ☆ Sensor Response Time Test (Optional) |
|
| ☆ Current Module Combined Output |
|
| ☆Current Supports Program-Controlled Commutation |
|
| ☆ Direct Display Of Sensor Measurement Errors |
|
| Current Range | 1 A, 2 A, 5 A, 10 A, 20 A, 50 A, 100 A, 200 A, 500 A….N kA (Optional) | ||
| Output Range | (10%~120%)*RG, using relay program-controlled commutation | ||
| Accuracy | Class 0.01 | Class 0.02 | Class 0.05 |
| Short Term Stability (%*RG/min) |
0.003 | 0.005 | 0.01 |
| Measurement Uncertainty (k=2) (%*RD+%*RG) [1] |
0.006 + 0.004 | 0.012 + 0.008 | 0.03 + 0.02 |
| Display Digits | 7-digit decimal | 7-digit decimal | 6-digit decimal |
| Adjust Fineness | 0.001%*RG | 0.002%*RG | 0.002%*RG |
| Maximum Load Voltage | 3.5 V | ||
| Ripple Coefficient | ≤ 0.5% | ||
| Setup Time | ≤ 3 s | ||
| Adjustment Method | Adjust the output current value through mobile measurement and control console and host computer software | ||
| Protective Function | Open circuit protection, overload protection, overheating protection | ||
| Note | [1]: RD is the reading, RG is the range, the same below. | ||
| Sensor Secondary Signal Measurement |
Voltage measurement range | 100 mV, 1 V, 10 V, manual or automatic shifting |
| ±(10mV~12V) | ||
| Current measurement range | 10 mA, 100 mA, 1 A, manual or automatic shifting | |
| ±(1 mA ~ 1.1 A) | ||
| Measurement uncertainty (k=2) | 0.003%*RD + 0.002%*RG | |
| display digits | 7-digit decimal | |
| Temperature Coefficient | 5 ppm/°C @ (0°C~40°C) | |
| Sensor Power Supply | Supply voltage | DC ±(5.0 V~50.0 V) adjustable |
| Maximum load capacity | 1A | |
| Measurement uncertainty (k=2) | Voltage/Current: 0.2%, Power: 0.5% | |
| Protective function | Short circuit protection, overload protection, overheating protection | |
| AC power supply (customized) | AC 220 V power supply can be added according to user needs | |
| Note | Sensor units can be customized according to user needs | |
| Current Output Range | 1mA~10.5A |
| Frequency Range | DC, 10 Hz ~ 1 MHz |
| Short Term Stability | DC: 0.005%/min |
| Measurement Uncertainty (k=2) |
AC: 0.01%/min @ 10 kHz, 0.05%/min @ 100kHz, 0.1%/min@1MHz |
| Max Load Voltage | DC: 0.02% |
| Application | AC: 0.08% @ 10 kHz, 0.5% @ 100 kHz |
| Current Output Range | 500A |
| Pulse type | Unipolar pulse |
| Measurement Uncertainty (k=2) | 0.5% |
| Pulse Width | 1 ms ~ 100 ms, adjustment step 0.1 ms |
| Rise/fall Rate | >50 A/μs |
| Max Load Voltage | 9 Vpk |
| Application | Current sensor response time test |
| Nominal Input Current | Two specifications: 100 mA and 1 A |
| Nominal Output Voltage | 1 V |
| Measurement Frequency Range | DC~1MHz |
| Resistance Annual Stability | 18 ppm |
| AC/DC Difference | 20 ppm @ 50 Hz, 25 ppm @ 100 kHz |
| Phase Difference | 5 μrad @ 50 Hz, 500 μrad @ 100 kHz |
| Application | Convert secondary current signals from Type I/I sensors to voltage for bandwidth or response time testing |
| Power Supply | Three-phase five-wire, AC (380±38) V, (50±2) Hz |
| Maximum Power Consumption | Maximum power consumption is 4 kVA per 500 A current source |
| Preheat Time | 30 minutes |
| Working Environment | Temperature: 0°C ~ 40°C Humidity: 20%R·H ~ 85%R·H, no condensing. Others: No electromagnetic field interference |
| Storage Environment | Temperature: -20°C ~ 70°C Humidity: 10%R·H ~ 95%R·H, no condensing. |
| Communication Interface | RS232×1 |