Customization: | Available |
---|---|
After-sales Service: | Life Time |
Warranty: | 12 Months Excellent After-Sales Service |
Shipping Cost: | Contact the supplier about freight and estimated delivery time. |
---|
Payment Methods: |
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() |
---|---|
Support payments in USD |
Secure payments: | Every payment you make on Made-in-China.com is protected by the platform. |
---|
Refund policy: | Claim a refund if your order doesn't ship, is missing, or arrives with product issues. |
---|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
☆ Set up AC and DC high voltage measurement and control system |
![]() |
|
☆ Calibrate AC and DC high Voltage Source / Meter |
![]() |
The secondary output voltage of TH0180 can be input to a high-precision voltmeter to achieve precision conversion measurement of AC and DC high voltage, which is used to measure the AC and DC high voltage of the power supply system, or for use Calibrate AC and DC high voltage sources such as withstand voltage testers, AC and DC high voltmeters and other equipment. |
☆ Calibrate AC/DC high-voltage V/V conversion equipment |
![]() |
TH0180 can be used as a standard to verify or calibrate AC and DC high-voltage dividers, transformers, sensors, high-voltage probes, etc. V/ by voltage ratio method or difference method (need to be the same ratio) V-conversion device to achieve calibration of its divider ratio. |
☆Set up AC and DC high voltage and large current power measurement system |
![]() |
TH0180 can be combined with high-current I/V or I/I conversion equipment (typical example: TH0500), standard power meter (typical example: TD1310 AC-DC standard meter, TH2000 vector voltage analyzer) to form a precision AC-DC high-voltage high-current power/electric energy measurement system. It is used to measure the output power of high-power AC/DC power supply, conversion efficiency of high-power conversion equipment, high-voltage current standard power source, and standard power meter. |
Specifications | Nominal input voltage ( kV ) |
Maximum input voltage ( kV ) |
Nominal output voltage ( V ) |
Proportional uncertainty ( k=2, ppm ) |
Temperature coefficient ( ppm/K ) |
TH0180-1k | 1 | 1.1 | 5 | 50 | 2 |
TH0180-2k | 2 | 2.2 | 5 | 50 | 2 |
TH0180-5k | 5 | 5.5 | 5 | 50 | 5 |
TH0180-10k | 10 | 11 | 5 | 50 | 5 |
TH0180-35k | 35 | 38.5 | 5 | 50 | 5 |
Specifications | AC/DC differential (ppm). | Phase shift ( μrad ). | ||||
53 Hz | 1 kHz | 10 kHz | 53 Hz | 1 kHz | 10 kHz | |
TH0180-1k | 10 | 100 | 1000 | 50 | 100 | 1000 |
TH0180-2k | 10 | 100 | 1000 | 50 | 120 | 1200 |
TH0180-5k | 10 | 100 | 1000 | 50 | 200 | 2000 |
TH0180-10k | 50 | 200 | - | 200 | 1500 | - |
TH0180-35k | 50 | 200 | - | 200 | 1500 | - |
Power supply | -- |
Warm-up time | 30 minutes |
Temperature performance | Working temperature: 5 °C ~ 40 °C; Storage temperature: -10 °C ~ 5 5 °C Note: When stored in harsh temperature environments, it may cause a temporary change in the resistance value and return to the original resistance value within the next 30 days. |
Humidity performance | Operating humidity: ( 30% ~ 70% ) R· H, no condensation Storage humidity: ( 20% ~ 80% ) R· H, no condensation |
Altitude | < 3000 m |
Connection method | Voltage input: ET-G 9 high voltage socket; Voltage output: Type-N male.![]() |
Specifications | Size ( mm ) ( W * D * H ) (feet and handles are not included) |
Weight (kg) |
Dimensional drawing |
TH0180-1k TH0180-2k TH0180-5k TH0180-10k |
450 × 450 × 540 | 22.6 | ![]() |
TH0180-35k | 1095 × 1095 × 1108 | 45.2 | ![]() |