☆ Set Up the Automatic Test System for Precision Wideband Current Source. |
|
Range | Nominal Resistance( Ω ) | Initial Deviation[1] ( ± μΩ/Ω ) |
Year Stability ( ± μΩ/Ω ) |
Temperature Coefficient ( ± μΩ/Ω/K ) |
Power Factor [2] ( ± μΩ/Ω ) |
1 mA | 1000 | 50 | 18 | 2.5 | 1 |
10 mA | 100 | 50 | 18 | 2.5 | 1 |
100 mA | 10 | 30 | 12 | 2.0 | 1 |
1 A | 1 | 30 | 12 | 2.0 | 3 |
10 A | 0.1 | 30 | 12 | 2.5 | 5 |
20 A | 0.05 | 30 | 12 | 2.5 | 8 |
Note[1]:The initial deviation refers to the deviation between the measured resistance value and the nominal resistance value after the rated DC current is applied for 10 minutes under the calibration temperature environment; [2]:The initial deviation of the DC resistance has included the power factor influence error under the nominal current. When measuring other non-nominal current values, additional measurement uncertainty should be introduced. The calculation formula is: |
Range | Primary Nominal Current ( A ) |
Secondary Nominal Voltage ( V ) |
AC/DC Difference[3] ( ppm ) | Maximum Working Current( A ) | Instant Current ( A, <5 s ) |
||
50 Hz | 1 kHz | 100 kHz | |||||
1 mA | 1 m | 1 | 25 | 50 | 150 | 2 m | 3 m |
10 mA | 10 m | 1 | 20 | 25 | 50 | 20 m | 150 m |
100 mA | 100 m | 1 | 20 | 25 | 25 | 200 m | 1.5 |
1 A | 1 | 1 | 20 | 25 | 30 | 1.5 | 4 |
10 A | 10 | 1 | 20 | 30 | 100 | 11 | 30 |
20 A | 20 | 1 | 20 | 30 | 100 | 22 | 50 |
Note[3]:The measured value of AC current should be calculated according to the following formula, and the unit of AC/ DC difference is ppm: |
Current Range | Phase Displacement ( μrad ) | |||||
53 Hz | 400 Hz | 1 kHz | 10 kHz | 50 kHz | 100 kHz | |
1 mA | 10 | 10 | 15 | 100 | 500 | 1000 |
10 mA | 5 | 5 | 10 | 100 | 400 | 800 |
100 mA | 5 | 5 | 10 | 30 | 150 | 300 |
1 A | 5 | 5 | 10 | 30 | 150 | 300 |
10 A | 10 | 20 | 40 | 150 | 800 | 1600 |
20 A | 10 | 25 | 50 | 200 | 1000 | 2000 |
Power Supply | AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz; |
Warm-up Time | Not less than 1 hour, and the reheating time after shutdown halfway should not be less than twice the shutdown time; |
Temperature Performance | Working temperature: 13 °C ~ 33 °C; Calibration temperature: 18 °C ~ 28 °C; Storage temperature: 5 °C ~ 45 °C Note: When stored in a harsh temperature environment, it may cause a temporary change in the resistance value and return to the original resistance value within the next 30 days. |
Humidity Performance | Working humidity: < 50% R·H Storage humidity: (15%~80%) R·H, non-condensing |
Altitude | < 3000 m |
Connection Method | Input connection: Type-N female plug; Output connection: Type-N female plug, banana socket; |