TD3810 Three-phase Energy Device Field Testing System
1. Summary
TD3810 is a comprehensive testing system dedicated to the on-site assessment of the AC energy meter verification device. The functional integration of the instrument is extremely high, and a set of equipment can complete the verification items specified by JJG 597 (insulation and voltage test require the cooperation of safety equipment). It is not only suitable for assessing the three-phase energy standard device, but also used as a high precision energy standard meter.
2. Features
- Power / power accuracy 0.01,0.02 optional.
- Voltage measurement range: 6 V ~528 V (customized for wider range).
- Current measurement range: 0.2 mA ~ 120 A.
- Basal wave frequency: 45 Hz ~65 Hz (optional 400Hz).
- The current is input by a pair of terminal posts without changing change.
- Phase measurement uncertainty up to 0.003° (class 0)..01
- Voltage and current support the fully automatic range shift.
- Both voltage and current support 2 to 127 harmonic measurements.
- It has the function of comprehensive statistical analysis of the measured electricity quantity.
- Support the graphical display of phor map, spectrum map and trend map.
- With the standard power pulse input / output function.
- In the limit state, the instrument is not damaged and the value is accurate.
- Support for temperature and humidity, magnetic field, daily timing error and other measurement functions
- Rich communication interfaces: USB, RS232, LAN.
- With a large size LCD touch touch screen.
3. Applications
☆ Assessment of Single-phase / Three-phase Energy Meter verification Device |
|
- On-site assessment of single phase electric energy meter verification device and single phase standard electricity meter with class 0.02 and below meeting R46 instruction
- The possible verification items are as follows:
|
Intrinsic Error |
Power Stability |
Interrelationship |
The Influence of Alternating Alternating Magnetic Field |
Multiplex Output Consistency |
Adjustable Range |
Measurement Repeatability |
Degree of Symmetry |
Potential Difference between the Terminals of the Same Name |
Short-term Stability Variation |
Fineness Adjustment |
The Magnetic Field of the Device |
Phase Sequence |
Monitoring Indication Error |
Variation Within Calibration Cycle |
Load Effect |
Waveform Distortion |
|
|
|
☆ Calibrate Single-phase / Three-phase Power Source / Meter |
|
- Suitable for calibrating single / three-phase power standard source and power meter of class 0.02 and below (with stable power source).
|
4.Characteristics
☆ Front Panel of the Instrument |
|
S/N |
Function |
1 |
Built handle, that is, does not occupy space and convenient for users to carry instruments. |
2 |
10.4-inch LCD touch color screen, multi-power intuitive display, full touch screen operation, more convenient operation. |
3 |
The folding foot can raise the instrument by about 15° for optimal operation and reading view. |
5.Functional characteristics
☆ Wide Range of Current Measurement |
Each phase is equipped with a pair of fast terminals and wires, and the ac current of 0.2 mA ~120 A can be measured by direct connection, simplifying the wiring operation. Fully automatic switching range, the transformer does not shift at one time, no matter in which range suddenly into the large current will not lead to damage or measurement accuracy reduction. The minimum measurement is limited to 0.2 mA, which can assess the accuracy and stability of the power / electric power of the R46 electric power device and its standard meter at the minimum starting current of 0.3 mA. |
☆ High Reliability of the Instrument |
There is complete electrical isolation between the voltage and current measurement circuits; in the limit state (500 V, 100 A), it will not damage the instrument, and can ensure the accurate value.
|
☆ Easy to Use On-site |
The main weight of the instrument is 9.9 kg and equipped with a special pull rod instrument box, which can put the host, test wire and laptop computer in it, and the transportation is very convenient. The notebook is equipped with special software meeting the requirements of JJG 597, which can automatically complete the error analysis, data recording, export or printing certificate of the device on site. |
6. Technical Specifications
6.1 Three-phase Voltage Measurement
Range |
Resolving Power |
Measurement Uncertainty (k = 2) ( ppm*RD + ppm*RG )[1] |
Temperature Coefficient @ (15~30) °C (ppm*RD+ppm *RG ) /°C |
Class 0.02 |
Class 0.01 |
Class 0.02 |
Class 0.01 |
60 V |
10 μV |
60 + 40 |
30 + 20 |
0.5 + 0.5 |
0.25 + 0.25 |
120 V |
0.1mV |
60 + 40 |
30 + 20 |
0.5 + 0.5 |
0.25 + 0.25 |
240 V |
0.1mV |
60 + 40 |
30 + 20 |
0.5 + 0.5 |
0.25 + 0.25 |
480 V |
0.1mV |
60 + 40 |
30 + 20 |
0.5 + 0.5 |
0.25 + 0.25 |
Note {1]: RD is the read value and RG is the range value, the same below. |
- Measurement range: 6 V ~528 V (wider range supports customization), 7-bit display, manual / automatic range shift
6.2 Three-phase Current Measurement
Class 0.01 Specification |
Range |
Resolving Power |
Measurement Uncertainty at different frequencies (Hz) (k = 2) ( ppm*RD + ppm*RG ) |
Temperature Coefficient @ (15~30) °C (ppm*RD+ppm *RG) /°C |
45 ≤ F ≤ 65 |
65 < F ≤ 200 |
200 < F ≤ 400 |
5 mA |
1 nA |
120 + 80 |
240 + 160 |
480 + 320 |
5 + 5 |
10 mA |
10 nA |
60 + 40 |
120 + 80 |
240 + 160 |
3 + 3 |
20 mA |
10 nA |
60 + 40 |
120 + 80 |
240 + 160 |
0.75 + 0.75 |
50 mA |
10 nA |
30 + 20 |
60 + 40 |
120 + 80 |
0.25 + 0.25 |
100 mA |
0.1 μA |
30 + 20 |
60 + 40 |
120 + 80 |
0.25 + 0.25 |
200 mA |
0.1 μA |
30 + 20 |
60 + 40 |
120 + 80 |
0.25 + 0.25 |
500 mA |
0.1 μA |
30 + 20 |
60 + 40 |
120 + 80 |
0.25 + 0.25 |
1 A |
1 μA |
30 + 20 |
60 + 40 |
120 + 80 |
0.25 + 0.25 |
2 A |
1 μA |
30 + 20 |
60 + 40 |
120 + 80 |
0.25 + 0.25 |
5 A |
1 μA |
30 + 20 |
60 + 40 |
120 + 80 |
0.25 + 0.25 |
10 A |
10 μA |
30 + 20 |
60 + 40 |
120 + 80 |
0.25 + 0.25 |
20 A |
10 μA |
30 + 20 |
60 + 40 |
120 + 80 |
0.25 + 0.25 |
50 A |
10 μA |
30 + 20 |
60 + 40 |
120 + 80 |
0.25 + 0.25 |
100 A |
100 μA |
30 + 20 |
60 + 40 |
120 + 80 |
0.25 + 0.25 |
- Measurement range: 0.2 mA ~ 120 A, 7-bit display, manual / automatic range shift
Class 0.02 Specification |
Range |
Resolving Power |
Measurement Uncertainty at different frequencies (Hz) (k = 2) ( ppm*RD + ppm*RG ) |
Temperature Coefficient @ (15~30) °C (ppm*RD+ppm *RG) /°C |
45 ≤ F ≤ 65 |
65 < F ≤ 200 |
200 < F ≤ 400 |
5 mA |
1 nA |
240 + 160 |
480 + 320 |
1000 + 600 |
10 + 10 |
10 mA |
10 nA |
120 + 80 |
240 + 160 |
480 + 320 |
5 + 5 |
20 mA |
10 nA |
120 + 80 |
240 + 160 |
480 + 320 |
1.25 + 1.25 |
50 mA |
10 nA |
60 + 40 |
120 + 80 |
240 + 160 |
0.5 + 0.5 |
100 mA |
0.1 μA |
60 + 40 |
120 + 80 |
240 + 160 |
0.5 + 0.5 |
200 mA |
0.1 μA |
60 + 40 |
120 + 80 |
240 + 160 |
0.5 + 0.5 |
500 mA |
0.1 μA |
60 + 40 |
120 + 80 |
240 + 160 |
0.5 + 0.5 |
1 A |
1 μA |
60 + 40 |
120 + 80 |
240 + 160 |
0.5 + 0.5 |
2 A |
1 μA |
60 + 40 |
120 + 80 |
240 + 160 |
0.5 + 0.5 |
5 A |
1 μA |
60 + 40 |
120 + 80 |
240 + 160 |
0.5 + 0.5 |
10 A |
10 μA |
60 + 40 |
120 + 80 |
240 + 160 |
0.5 + 0.5 |
20 A |
10 μA |
60 + 40 |
120 + 80 |
240 + 160 |
0.5 + 0.5 |
50 A |
10 μA |
60 + 40 |
120 + 80 |
240 + 160 |
0.5 + 0.5 |
100 A |
100 μA |
60 + 40 |
120 + 80 |
240 + 160 |
0.5 + 0.5 |
- Measurement range: 0.2 mA ~ 120 A, 7-bit display, manual / automatic range shift
6.3 Frequency / Phase Measurement
Class 0.01 Specification |
Type of Measurement |
TD3810 |
TD3810-R |
Frequency |
measuring range |
45 Hz ~65 Hz |
45 Hz ~400 Hz |
Minimum resolution |
0.000 01 Hz |
0.000 01 Hz |
Measurement uncertainty (k=2) |
0.005%*RD |
0.005%*RD |
Phase Position |
measuring range |
0~360°(I ≥ 50mA ) |
0~360°(I ≥ 50mA ) |
Minimum resolution |
0.000 1° |
0.000 1° |
Measurement uncertainty (k=2) |
45 Hz ≤ F ≤ 65 Hz |
0.003° |
0.003° |
65 Hz < F ≤ 200 Hz |
-- |
0.01° |
200 Hz < F ≤ 400 Hz |
-- |
0.02° |
Class 0.02 Specification |
Type of Measurement |
TD3810 |
TD3810-R |
Frequency |
measuring range |
45 Hz ~65 Hz |
45 Hz ~400 Hz |
Minimum resolution |
0.000 01 Hz |
0.000 01 Hz |
Measurement uncertainty (k=2) |
0.005%*RD |
0.005%*RD |
Phase Position |
measuring range |
0~360°(I ≥ 50mA ) |
0~360°(I ≥ 50mA ) |
Minimum resolution |
0.000 1° |
0.000 1° |
Measurement uncertainty (k=2) |
45 Hz ≤ F ≤ 65 Hz |
0.006° |
0.006° |
65 Hz < F ≤ 200 Hz |
-- |
0.02° |
200 Hz < F ≤ 400 Hz |
-- |
0.04° |
6.4 Power / Electric Energy Measurement
Class 0.02 Specification |
Voltage Range |
Current range |
Power Factor |
Measurement uncertainty at different frequencies (Hz) (k=2) |
45 ≤ F ≤ 65 |
65 < F ≤ 200 |
200 < F ≤ 400 |
30 V ≤ U ≤ 480 V |
50 mA ≤ I ≤120 A |
0.5L~1~0.5C |
0.01%*RD |
0.02%*RD |
0.04%*RD |
10 mA ≤ I < 50 mA |
1 |
0.01%*RD |
0.03%*RD |
0.08%*RD |
0.5L~1~0.5C |
0.02%*RD |
3 mA ≤ I < 10 mA |
1 |
0.02%*RD |
- |
- |
0.5L~1~0.5C |
0.04%*RD |
- |
- |
0.2 mA ≤ I < 3 mA |
1 |
0.02%*RD×3mA/I |
- |
- |
Class 0.05 Specification |
Voltage Range |
Current Range |
Power Factor |
Measurement Uncertainty at different frequencies (Hz) (k=2) |
45 ≤ F ≤ 65 |
65 < F ≤ 200 |
200 < F ≤ 400 |
30 V ≤ U ≤ 480 V |
50 mA ≤ I ≤120 A |
0.5L~1~0.5C |
0.02%*RD |
0.04%*RD |
0.08%*RD |
10 mA ≤ I < 50 mA |
1 |
0.02%*RD |
0.06%*RD |
0.16%*RD |
0.5L~1~0.5C |
0.04%*RD |
3 mA ≤ I < 10 mA |
1 |
0.04%*RD |
- |
- |
0.5L~1~0.5C |
0.08%*RD |
- |
- |
0.2 mA ≤ I < 3 mA |
1 |
0.04%*RD×3mA/I |
- |
- |
- Power / electric power measurement range: the combination of AC voltage range and AC flow range
- Power factor measurement range: -1.000 000... 0.000 000... 1.000 000
- Standard electric energy pulse output: high frequency full range value corresponds to 60 kHz, low frequency full range value corresponds to 6 Hz
- Standard electric energy pulse input: frequency 200 kHz, voltage: 0... 3 V... 24 V
6.5 Other Measurement Functions
Device Magnetic Field Measurement |
Range |
100 μ T, 1 mT, and 5 bits are shown |
Measurement Uncertainty (k=2) |
0.5% |
Probe |
Three-dimensional magnetic field sensing probe |
Remarks |
To measure the position of the measuring meter of the electric energy meter verification device |
Object Difference Measurement |
Range |
5 mV ,50 mV ,500 mV ,5V |
Measurement Uncertainty (k=2) |
0.2 mV |
Remarks |
Can to withstand 250 V |
Clock Measurement |
Function |
PPS standard second pulse output, measurement |
Measurement Uncertainty (k=2) |
5×10-8 |
7. Ordering Information