Customization: | Available |
---|---|
After-sales Service: | Life Time |
Warranty: | 12 Months Excellent After-sales Service |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Range | Short-Term Stability (%/Min) |
Optimum Measurement Uncertainty (k=2) (Ppm*Rd+Ppm *Rg) |
Maximum Load Current (mA) |
||||
Class 0.05 |
Class 0.02 |
Class 0.01 |
Class 0.05 |
Class 0.02 |
Class 0.01 |
||
10 mV | 0.01 | 0.005 | 0.003 | 200 + 5 μV | 100 + 5 μV | 100 + 5 μV | 100 |
30 mV | 0.01 | 0.005 | 0.003 | 200 + 5 μV | 100 + 5 μV | 100 + 5 μV | 100 |
100 mV | 0.01 | 0.005 | 0.003 | 200 + 5 μV | 100 + 5 μV | 100 + 5 μV | 100 |
300 mV | 0.01 | 0.005 | 0.003 | 200 + 5 μV | 100 + 5 μV | 100 + 5 μV | 100 |
1 V | 0.005 | 0.005 | 0.002 | 200 + 100 | 100 + 50 | 60 + 40 | 500 |
3 V | 0.005 | 0.005 | 0.002 | 200 + 100 | 100 + 50 | 60 + 40 | 500 |
10 V | 0.005 | 0.005 | 0.002 | 200 + 100 | 100 + 50 | 60 + 40 | 500 |
30 V | 0.005 | 0.005 | 0.002 | 300 + 200 | 120 + 80 | 60 + 40 | 800 |
100 V | 0.005 | 0.005 | 0.002 | 300 + 200 | 120 + 80 | 60 + 40 | 300 |
300 V | 0.005 | 0.005 | 0.002 | 300 + 200 | 120 + 80 | 60 + 40 | 100 |
600 V | 0.005 | 0.005 | 0.002 | 300 + 200 | 120 + 80 | 60 + 40 | 50 |
1000 V | 0.005 | 0.005 | 0.002 | 300 + 200 | 120 + 80 | 60 + 40 | 30 |
Note : RD is the read value and RG is the range value |
Range |
Short-Term Stability (%/Min) |
Optimum Measurement Uncertainty (k=2) (Ppm*Rd+Ppm *Rg) | Maximum Load Voltage (V) |
||||
Class 0.05 | Class 0.02 | Class 0.01 | Class 0.05 | Class 0.02 | Class 0.01 | ||
10 μA | 0.01 | 0.01 | 0.01 | 300 + 200 | 300 + 200 | 300 + 200 | 1 1 |
30 μA | 0.01 | 0.005 | 0.005 | 300 + 200 | 120 + 80 | 120 + 80 | 1 1 |
100 μA | 0.01 | 0.005 | 0.005 | 300 + 200 | 120 + 80 | 120 + 80 | 1 1 |
300 μA | 0.01 | 0.005 | 0.005 | 300 + 200 | 120 + 80 | 120 + 80 | 1 1 |
1 mA | 0.01 | 0.005 | 0.003 | 300 + 200 | 120 + 80 | 100 + 50 | 1 1 |
3 mA | 0.01 | 0.005 | 0.003 | 300 + 200 | 120 + 80 | 100 + 50 | 1 1 |
10 mA | 0.01 | 0.005 | 0.003 | 300 + 200 | 120 + 80 | 100 + 50 | 1 1 |
30 mA | 0.01 | 0.005 | 0.003 | 300 + 200 | 120 + 80 | 100 + 50 | 1 1 |
100 mA | 0.01 | 0.005 | 0.003 | 300 + 200 | 120 + 80 | 100 + 50 | 1 1 |
300 mA | 0.01 | 0.005 | 0.002 | 300 + 200 | 120 + 80 | 60 + 40 | 1 1 |
1 A | 0.01 | 0.005 | 0.002 | 300 + 200 | 120 + 80 | 60 + 40 | 1 1 |
3 A | 0.01 | 0.005 | 0.002 | 300 + 200 | 120 + 80 | 60 + 40 | 4 |
10 A | 0.01 | 0.005 | 0.002 | 300 + 200 | 120 + 80 | 60 + 40 | 4 |
30 A | 0.01 | 0.005 | 0.002 | 300 + 200 | 120 + 80 | 60 + 40 | 3 |
100 A | 0.01 | 0.005 | 0.002 | 300 + 200 | 120 + 80 | 60 + 40 | 3 |
Dc Power / Power |
Output Range |
A combination of a DC voltage and a DC current (or a small DC signal voltage) |
Uncertainty Of Measurement |
Voltage measurement uncertainty + uncertainty of current measurement (or uncertainty of small DC signal voltage measurement) |
|
Electric Energy Pulse |
Standard Electrical Energy Pulse Input |
The high-frequency full range value corresponds to 60 kHz; The low-frequency full range value corresponds to 6 Hz; |
Pulse Output Frequency |
The output frequency jitter time was < 10 μ s |
|
Annual Stability Of The Clock Frequency |
Better than ± 50 ppm |
|
Standard Electrical Energy Pulse Input | Signal amplitude is 5V, with a TTL Class; capable ≤ 200 kHz; FH = 60 kHz power value / voltage range value / current range value FL = 60 kHz power value / voltage range value / current range value / 10000 |
|
Electric Energy Error Display |
Automatically displayed with a resolution of 0.0001% |
Range |
Short-term Stability (%/min) |
Measurement Uncertainty (k=2) (ppm*RD+ppm *RG/μV ) |
Maximum Load Current (mA) |
1 mV | 1 μV | 120 + 2 μV | ≤ 10 |
3 mV | 1 μV | 120 + 2 μV | ≤ 10 |
10 mV | 1 μV | 120 + 2 μV | ≤ 10 |
30 mV | 1 μV | 120 + 2 μV | ≤ 10 |
100 mV | 0.005 | 120 + 80 | ≤ 10 |
300 mV | 0.005 | 120 + 80 | ≤ 10 |
1 V | 0.005 | 120 + 80 | ≤ 10 |
4 V | 0.005 | 120 + 80 | ≤ 10 |
Voltage / Current range | Measuring Range | Optimum Measurement Uncertainty (k=2) (ppm*RD+ppm *RG) |
1 V | ±(0~ 1.2) V | 60 + 40 |
10 V | ±(0~ 12) V | 60 + 40 |
2 mA | ±(0~2.4) mA | 60 + 40 |
20 mA | ±(0~24) mA | 60 + 40 |
Note : RD is the reading value and RG is the range value |
Power Supply | AC (220±22) V ,(50±2) Hz |
Preheating time | 30 Minutes |
Max. Power dissipation |
700 VA |
Temperature Performance |
Operating temperature: 0°C~40°C; Calibration temperature: 18°C~28°C; Storage temperature: -20°C~70°C |
Humidity Performance |
Working Humidity: <80% @ 30°C, <70% @ 40°C, and <40% @ 50°C Storage humidity: (20%~80%) R · H, no condensation |
Altitude | < 3000 m |
Weight | About 17.5kg |
Connect mode | The RS232 interface |
Outline | 446 mm (W) × 522 mm (D) × 197 mm (H) |